Index: firmware/App/Modes/FPModes/InletPressureCheck.c =================================================================== diff -u -rd25471a9b256c1dce1aec7de73f53c730ff70cd0 -rdd0e9acd5f0fb570403be8908ac45e79840d4b70 --- firmware/App/Modes/FPModes/InletPressureCheck.c (.../InletPressureCheck.c) (revision d25471a9b256c1dce1aec7de73f53c730ff70cd0) +++ firmware/App/Modes/FPModes/InletPressureCheck.c (.../InletPressureCheck.c) (revision dd0e9acd5f0fb570403be8908ac45e79840d4b70) @@ -158,22 +158,22 @@ } } // reset the persistence timer when inlet pressure out of range - else if( ( inletPressure <= thresholdMin ) || ( inletPressure >= thresholdMax ) ) + else if ( ( inletPressure <= thresholdMin ) || ( inletPressure >= thresholdMax ) ) { inletPressureCheckPersistanceStartTimeMS = 0; // Trigger alarm if M3 pressure is not within range if ( TRUE == didTimeout( inletPressureCheckTimer, getInletPressureCheckTimeout() ) ) { - SET_ALARM_WITH_2_U32_DATA( ALARM_ID_FP_RO_INLET_PRESSURE_OUT_HIGH_RANGE, INLET_PRES_CHECK_TIMEOUT, inletPressure) + SET_ALARM_WITH_2_U32_DATA( ALARM_ID_FP_RO_INLET_PRESSURE_OUT_HIGH_RANGE, INLET_PRES_CHECK_TIMEOUT, inletPressure ) state = INLET_PRES_CHECK_PAUSED; } } } else { - if( inletPressure >= INLET_PRES_CHECK_TGT_PSI ) + if ( inletPressure >= INLET_PRES_CHECK_TGT_PSI ) { - SET_ALARM_WITH_2_U32_DATA( ALARM_ID_FP_RO_INLET_PRESSURE_OUT_HIGH_RANGE, INLET_PRES_CHECK_TGT_PSI, inletPressure) + SET_ALARM_WITH_2_U32_DATA( ALARM_ID_FP_RO_INLET_PRESSURE_OUT_HIGH_RANGE, INLET_PRES_CHECK_TGT_PSI, inletPressure ) state = INLET_PRES_CHECK_PAUSED; } else