Index: firmware/App/Modes/FPModes/StateFlushPermeate.c =================================================================== diff -u -r9f1f5a95111b55ee99bbdcd60a2c8b4cc088237c -r4626638b758c6673c0a8a3cc8d008bb4a898cea9 --- firmware/App/Modes/FPModes/StateFlushPermeate.c (.../StateFlushPermeate.c) (revision 9f1f5a95111b55ee99bbdcd60a2c8b4cc088237c) +++ firmware/App/Modes/FPModes/StateFlushPermeate.c (.../StateFlushPermeate.c) (revision 4626638b758c6673c0a8a3cc8d008bb4a898cea9) @@ -41,9 +41,9 @@ #define PERMEATE_FLUSH_TIMEOUT ( 2 * SEC_PER_MIN * MS_PER_SECOND ) ///< Permeate flush timer (in ms) timeout for 2 minutes #define PERMEATE_FLUSH_ALARM_TIMEOUT ( 5 * SEC_PER_MIN * MS_PER_SECOND ) ///< Permeate flush alarm timer (in ms) timeout for 5 minutes #define PERMEATE_FLUSH_RO_PUMP_TGT_ML 800 ///< Pressure target in ml/min for the RO pump during permeate flush. -#define PERMEATE_FLUSH_BOOST_PUMP_TGT_PSI 25 ///< Pressure target in PSI for the boost pump during permeate flush. +#define PERMEATE_FLUSH_BOOST_PUMP_TGT_PSI 8 ///< Pressure target in PSI for the boost pump during permeate flush. #define PERMEATE_FLUSH_CONDUCTIVITY_THRESHOLD 200.0F ///< Conductivity alarm threshold for permeate flush. -#define MIN_BOOST_PUMP_DUTY_CYCLE_PCT 0.5F ///< Minimum duty cycle value for boost pump during open loop control +#define MIN_BOOST_PUMP_DUTY_CYCLE_PCT 0.1F ///< Minimum duty cycle value for boost pump during open loop control // ********** private data ********** @@ -262,7 +262,6 @@ { setBoostPumpTargetPressure( PERMEATE_FLUSH_BOOST_PUMP_TGT_PSI, TRUE ); } - //TODO update target macro after changing it to 750 + rinse pump rate setROPumpTargetFlowRateMLPM( PERMEATE_FLUSH_RO_PUMP_TGT_ML, TRUE ); permeateFlushTimer = getMSTimerCount(); permeateFlushAlarmTimer = getMSTimerCount();