Index: firmware/App/Modes/ModeGenDialysate.c =================================================================== diff -u -r5075786897d42d24a93ae5485936b5d50c307f65 -r3cafed12c435f95be8ae38a8e5a5cf20ec6b706f --- firmware/App/Modes/ModeGenDialysate.c (.../ModeGenDialysate.c) (revision 5075786897d42d24a93ae5485936b5d50c307f65) +++ firmware/App/Modes/ModeGenDialysate.c (.../ModeGenDialysate.c) (revision 3cafed12c435f95be8ae38a8e5a5cf20ec6b706f) @@ -72,7 +72,7 @@ //Testing #define DELAY_BC_SWITCHING_AT_START_UP ( 10 * MS_PER_SECOND ) ///< Provide a balancing chamber switching start up delay to stabilize pump speed etc., -#define MAX_UF_RATE 150 +#define MAX_UF_RATE 150 ///< Rate at which D76 pump runs in Prime fluid discard state /// Payload record structure for Gen dialysate execution state set request typedef struct @@ -491,8 +491,7 @@ break; case DD_GEND_PRIME_FLUID_DISCARD_STATE: - // Deactivate Balancing Chamber Switching - //requestBalChamberSwitching( FALSE ); + // Turn off fresh & spent pumps signalDialysatePumpHardStop( D12_PUMP ); signalDialysatePumpHardStop( D48_PUMP ); @@ -846,7 +845,6 @@ isDialDeliveryInProgress.data = FALSE; state = DD_GEND_PRIME_FLUID_DISCARD_STATE; } - } return state; @@ -983,7 +981,7 @@ * The handleGenDPrimeFluidDiscardState function performs discarding the * priming fluid from disposable tube set. Runs the ultrafiltration pump (D76) * to discard the priming fluid. - * @details \b Inputs: none + * @details \b Inputs: getprimeFluidDiscardStatus * @details \b Outputs: none * @return next gen dialysate mode state *************************************************************************/