Index: firmware/App/Modes/ModeGenDialysate.c =================================================================== diff -u -r937362196a40d7d55b1b61e33f88c382184fbf2f -rbd896114f304304a7096b30b2a85067a64645e82 --- firmware/App/Modes/ModeGenDialysate.c (.../ModeGenDialysate.c) (revision 937362196a40d7d55b1b61e33f88c382184fbf2f) +++ firmware/App/Modes/ModeGenDialysate.c (.../ModeGenDialysate.c) (revision bd896114f304304a7096b30b2a85067a64645e82) @@ -7,8 +7,8 @@ * * @file ModeGenDialysate.c * -* @author (last) “rkallala” -* @date (last) 27-Jan-2026 +* @author (last) Michael Garthwaite +* @date (last) 29-Jan-2026 * * @author (original) Vinayakam Mani * @date (original) 06-Nov-2024 @@ -961,7 +961,7 @@ *************************************************************************/ static void checkDialysateTemperature( void ) { - F32 dialysateTemp = getConductivityTemperatureValue( D29_COND ); // Assuming the closest temp sensor to dialyzer + F32 dialysateTemp = getFilteredConductivitySensorTemperature( D29_COND ); // Assuming the closest temp sensor to dialyzer F32 targetTemp = getTDTargetDialysateTemperature(); BOOL isDialTempAboveHighSafety = ( dialysateTemp >= DIALYSATE_TEMP_UPPER_MAX_SAFETY_LIMIT_C ? TRUE : FALSE ); BOOL isDialTempAboveLowSafety = ( dialysateTemp > DIALYSATE_TEMP_UPPER_SAFETY_LIMIT_C ? TRUE : FALSE );