Index: firmware/App/Modes/ModeGenDialysate.c =================================================================== diff -u -rf40a7a70ad79b1dfba9d3b19887326d0a923a775 -rc804832f7c982e331bf18a239e6815acf915c4c3 --- firmware/App/Modes/ModeGenDialysate.c (.../ModeGenDialysate.c) (revision f40a7a70ad79b1dfba9d3b19887326d0a923a775) +++ firmware/App/Modes/ModeGenDialysate.c (.../ModeGenDialysate.c) (revision c804832f7c982e331bf18a239e6815acf915c4c3) @@ -959,11 +959,7 @@ *************************************************************************/ static void checkDialysateTemperature( void ) { -#ifdef __TEENSY_CONDUCTIVITY_DRIVER__ - F32 dialysateTemp = getTeensyConductivityTemperatureValue( D29_COND ); // Assuming the closest temp sensor to dialyzer -#else - F32 dialysateTemp = getConductivityTemperatureValue( D29_COND ); // Assuming the closest temp sensor to dialyzer -#endif + F32 dialysateTemp = getFilteredConductivitySensorTemperature( D29_COND ); // Assuming the closest temp sensor to dialyzer F32 targetTemp = getTDTargetDialysateTemperature(); BOOL isDialTempAboveHighSafety = ( dialysateTemp >= DIALYSATE_TEMP_UPPER_MAX_SAFETY_LIMIT_C ? TRUE : FALSE ); BOOL isDialTempAboveLowSafety = ( dialysateTemp > DIALYSATE_TEMP_UPPER_SAFETY_LIMIT_C ? TRUE : FALSE );