Index: firmware/App/Modes/ModeGenDialysate.h =================================================================== diff -u -rf1dc3383b840c39f202b15679e0bbe2b2f7a2f83 -r847659965e5823e2e25db62c14f4611db9527f8a --- firmware/App/Modes/ModeGenDialysate.h (.../ModeGenDialysate.h) (revision f1dc3383b840c39f202b15679e0bbe2b2f7a2f83) +++ firmware/App/Modes/ModeGenDialysate.h (.../ModeGenDialysate.h) (revision 847659965e5823e2e25db62c14f4611db9527f8a) @@ -8,7 +8,7 @@ * @file ModeGenDialysate.h * * @author (last) Jashwant Gantyada -* @date (last) 07-May-2026 +* @date (last) 20-Jul-2026 * * @author (original) Vinayakam Mani * @date (original) 06-Nov-2024 @@ -32,7 +32,7 @@ // ********** public definitions ********** #define FRESH_DIAL_PUMP_INITIAL_RPM 2500 ///< Nominal RPM target for fresh dialysate pump to maintain required pressure. -#define FRESH_DIAL_PUMP_INITIAL_RPM_B1_9_B2_0 1600 ///< Nominal RPM target for fresh dialysate pump for beta 1.9 and 2.0 hardware. +#define FRESH_DIAL_PUMP_INITIAL_RPM_B1_9_B2_0 1250 ///< Nominal RPM target for fresh dialysate pump for beta 1.9 and 2.0 hardware. #define SPENT_DIAL_PUMP_INITIAL_RPM 2300 ///< Nominal RPM target for spent dialysate pump to maintain required pressure. #define DIAL_PUMP_DRAIN_RPM 1000 ///< Nominal RPM target for dialysate pump to drain the dry bicart. #define SPENT_DIAL_PUMP_FILL_RPM 200 ///< Nominal RPM target for spent chamber fill operations.