Index: firmware/App/Monitors/Conductivity.c =================================================================== diff -u -rf19fa057e6227ce04670a0976d121bb684e06e7e -rdeba34e5074da9ac97d8bcbe995f554eec5c70ed --- firmware/App/Monitors/Conductivity.c (.../Conductivity.c) (revision f19fa057e6227ce04670a0976d121bb684e06e7e) +++ firmware/App/Monitors/Conductivity.c (.../Conductivity.c) (revision deba34e5074da9ac97d8bcbe995f554eec5c70ed) @@ -35,8 +35,8 @@ #define COND_SENSOR_REPORT_PERIOD ( MS_PER_SECOND / TASK_PRIORITY_INTERVAL ) ///< Broadcast conductivity values message every second. #define COND_SENSOR_UPDATE_INTERVAL ( 700 / TASK_PRIORITY_INTERVAL ) ///< Time in task intervals for new sensor data #define DATA_PUBLISH_COUNTER_START_COUNT 40 ///< Data publish counter start count. -#define CONDUCTIVITY_SAMPLE_FILTER_MS ( 500 ) ///< Filter conductivity data for given time -#define CONDUCTIVITY_TEMP_SAMPLE_FILTER_MS ( 500 ) ///< Filter conductivity temperature data for given time +#define CONDUCTIVITY_SAMPLE_FILTER_MS ( 30 ) ///< Filter conductivity data for given time +#define CONDUCTIVITY_TEMP_SAMPLE_FILTER_MS ( 30 ) ///< Filter conductivity temperature data for given time #define SIZE_OF_COND_ROLLING_AVG ( CONDUCTIVITY_SAMPLE_FILTER_MS / TASK_PRIORITY_INTERVAL ) ///< Filtered conductivity moving average sample count. #define SIZE_OF_COND_TEMP_ROLLING_AVG ( CONDUCTIVITY_TEMP_SAMPLE_FILTER_MS / TASK_PRIORITY_INTERVAL ) ///< Filtered conductivity temprature moving average sample count. #define RO_RR_MOVING_AVG_NUM_OF_SAMPLES 30 ///< RO rejection ratio moving average number of samples.