Index: firmware/App/Monitors/Conductivity.c =================================================================== diff -u -rff3abe5d9c85e6c0431be7310ab771aebd9a8823 -rff6b765b44991e46891475022fa819cf3af05156 --- firmware/App/Monitors/Conductivity.c (.../Conductivity.c) (revision ff3abe5d9c85e6c0431be7310ab771aebd9a8823) +++ firmware/App/Monitors/Conductivity.c (.../Conductivity.c) (revision ff6b765b44991e46891475022fa819cf3af05156) @@ -38,7 +38,7 @@ #define FP_CONDUCTIVITY_DATA_PUBLISH_COUNTER_START_COUNT 41 ///< Data publish counter start count. #define RO_DATA_PUBLISH_COUNTER_START_COUNT 42 ///< Data publish counter start count. #define RESISTANCE_DATA_PUBLISH_COUNTER_START_COUNT 43 ///< Data publish counter start count. -#define CONDUCTIVITY_SAMPLE_FILTER_MS ( 30 ) ///< Filter conductivity data for given time. Currently set to have 5 samples over 3.5s ( 700ms sample rate ) +#define CONDUCTIVITY_SAMPLE_FILTER_MS ( 210 ) ///< Filter conductivity data for given time. Currently set to have 5 samples over 3.5s ( 700ms sample rate ) #define CONDUCTIVITY_TEMP_SAMPLE_FILTER_MS ( 30 ) ///< Filter conductivity temperature data for given time. Currently set to have 5 samples over 3.5s ( 700ms sample rate ) #define SIZE_OF_COND_ROLLING_AVG ( CONDUCTIVITY_SAMPLE_FILTER_MS / TASK_PRIORITY_INTERVAL ) ///< Filtered conductivity moving average sample count. #define SIZE_OF_COND_TEMP_ROLLING_AVG ( CONDUCTIVITY_TEMP_SAMPLE_FILTER_MS / TASK_PRIORITY_INTERVAL ) ///< Filtered conductivity temprature moving average sample count.