Index: firmware/App/Modes/FPModes/ModeGenPermeate.c =================================================================== diff -u -r3c6e233681b5a51632bdb6a06b20506fe50cd9f0 -r0778a3f68aec8f654ab001c8da3d447ccf86cf78 --- firmware/App/Modes/FPModes/ModeGenPermeate.c (.../ModeGenPermeate.c) (revision 3c6e233681b5a51632bdb6a06b20506fe50cd9f0) +++ firmware/App/Modes/FPModes/ModeGenPermeate.c (.../ModeGenPermeate.c) (revision 0778a3f68aec8f654ab001c8da3d447ccf86cf78) @@ -119,9 +119,6 @@ startPermeateTankControl(); setROPumpTargetFlowRateMLPM( GEN_PERMEATE_RO_PUMP_TGT_ML, TRUE ); - initPersistentAlarm( ALARM_ID_FP_PERMEATE_FLOW_OUT_LOW_RANGE, - 0, PERMEATE_FLOW_OUT_OF_RANGE_TIMEOUT_MS ); - return genPermeateState; } Index: firmware/App/Monitors/WaterQualityMonitor.c =================================================================== diff -u -r97f657d47a01b765b72e5193c434dd959c3a3f35 -r0778a3f68aec8f654ab001c8da3d447ccf86cf78 --- firmware/App/Monitors/WaterQualityMonitor.c (.../WaterQualityMonitor.c) (revision 97f657d47a01b765b72e5193c434dd959c3a3f35) +++ firmware/App/Monitors/WaterQualityMonitor.c (.../WaterQualityMonitor.c) (revision 0778a3f68aec8f654ab001c8da3d447ccf86cf78) @@ -94,6 +94,8 @@ //Flow Alarms initPersistentAlarm( ALARM_ID_FP_PERMEATE_FLOW_OUT_HIGH_RANGE, PERMEATE_FLOW_OUT_RANGE_PERSISTENCE_CLEAR_MS, PERMEATE_FLOW_OUT_RANGE_PERSISTENCE_TIMEOUT_MS ); + // fault , clearing timeout is not applicable + initPersistentAlarm( ALARM_ID_FP_PERMEATE_FLOW_OUT_LOW_RANGE, 0, PERMEATE_FLOW_OUT_OF_RANGE_TIMEOUT_MS ); } /*********************************************************************//**