Index: firmware/App/Controllers/ConcentratePumps.h =================================================================== diff -u -rf2b23abd44f71103b3ef6b0a96f18cc82d263735 -r352df4446493c8e071cc8e05fca8f191ac1bfba7 --- firmware/App/Controllers/ConcentratePumps.h (.../ConcentratePumps.h) (revision f2b23abd44f71103b3ef6b0a96f18cc82d263735) +++ firmware/App/Controllers/ConcentratePumps.h (.../ConcentratePumps.h) (revision 352df4446493c8e071cc8e05fca8f191ac1bfba7) @@ -33,7 +33,7 @@ #ifdef __DIENER_CONC_PUMP__ // For 100 RPM, Diener pump delivers 40ml/min,so restricing 300 RPM for now to maintain accuracy. We may allow up to 500 RPM // that yields 200 ml/min, but as speed increases, accuracy may drop. -#define CONCENTRATE_PUMP_MAX_SPEED 120.0F ///< Optimal speed for concentrate pump in mL/min +#define CONCENTRATE_PUMP_MAX_SPEED 60.0F ///< Optimal speed for concentrate pump in mL/min #define UF_PUMP_SPEED 100.0F ///< 100 ml/min, runs at 250 RPM #else #define CONCENTRATE_PUMP_MAX_SPEED 48.0F ///< Maximum speed for concentrate pump in mL/min Index: firmware/App/DDCommon.h =================================================================== diff -u -r51f20e0ccf76f2d178363bdf30a24127a75e20d1 -r352df4446493c8e071cc8e05fca8f191ac1bfba7 --- firmware/App/DDCommon.h (.../DDCommon.h) (revision 51f20e0ccf76f2d178363bdf30a24127a75e20d1) +++ firmware/App/DDCommon.h (.../DDCommon.h) (revision 352df4446493c8e071cc8e05fca8f191ac1bfba7) @@ -55,7 +55,7 @@ //#define ENABLE_ALARM_2 //Uncomment below if balancing chamber pressure alarm is required -#define __BC_PRESSURE_ALARM__ 1 +//#define __BC_PRESSURE_ALARM__ 1 //Uncomment below to disable heaters debug message #define __HEATERS_DEBUG__ 1