Index: firmware/App/Controllers/BalancingChamber.c =================================================================== diff -u -rbdee7caad13de484d50de4089497725c44a43c74 -r93282b8a03682015a0c70b94d95d387204e10a89 --- firmware/App/Controllers/BalancingChamber.c (.../BalancingChamber.c) (revision bdee7caad13de484d50de4089497725c44a43c74) +++ firmware/App/Controllers/BalancingChamber.c (.../BalancingChamber.c) (revision 93282b8a03682015a0c70b94d95d387204e10a89) @@ -529,7 +529,7 @@ // Check fresh and spent dialysate pressure in range or BC switch only flag set if ( ( ( freshDialPressure >= FRESH_DIAL_PRESSURE_MIN_PSIG ) && ( freshDialPressure <= FRESH_DIAL_PRESSURE_MAX_PSIG ) ) && ( ( spentDialPressure >= SPENT_DIAL_PRESSURE_MIN_PSIG ) && ( spentDialPressure <= SPENT_DIAL_PRESSURE_MAX_PSIG ) ) || - ( TRUE == getBalChamberSwitchingOnlyStatus() ) || ( TRUE == getTestConfigStatus( TEST_CONFIG_DISABLE_BC_PRESSURE_ALARMS ) ) ) + ( TRUE == getBalChamberSwitchingOnlyStatus() ) || ( TRUE == getTestConfigStatus( TEST_CONFIG_DD_DISABLE_BC_PRESSURE_ALARMS ) ) ) { //Valve control for state 1 fill valveControlForBCState1FillStart(); @@ -551,7 +551,7 @@ } else { - if ( getTestConfigStatus( TEST_CONFIG_DISABLE_BC_PRESSURE_ALARMS ) != TRUE ) + if ( getTestConfigStatus( TEST_CONFIG_DD_DISABLE_BC_PRESSURE_ALARMS ) != TRUE ) { //Alarm when pressure is not in range SET_ALARM_WITH_2_F32_DATA( ALARM_ID_DD_BC_STATE1_FILL_START_PRESSURE_OUT_OF_RANGE, freshDialPressure, spentDialPressure ); @@ -627,7 +627,7 @@ spentDialPressure = getFilteredPressure( D51_PRES ); // Check fresh and spent dialysate pressure back in range to indicate fill complete. - if ( getTestConfigStatus( TEST_CONFIG_DISABLE_BC_PRESSURE_ALARMS ) != TRUE ) + if ( getTestConfigStatus( TEST_CONFIG_DD_DISABLE_BC_PRESSURE_ALARMS ) != TRUE ) { if ( ( ( freshDialPressure >= FRESH_DIAL_PRESSURE_MIN_PSIG ) && ( freshDialPressure <= FRESH_DIAL_PRESSURE_MAX_PSIG ) ) && ( ( spentDialPressure >= SPENT_DIAL_PRESSURE_MIN_PSIG ) && ( spentDialPressure <= SPENT_DIAL_PRESSURE_MAX_PSIG ) ) ) @@ -683,7 +683,7 @@ if ( TRUE != getBalChamberSwitchingOnlyStatus() ) { - if ( getTestConfigStatus( TEST_CONFIG_DISABLE_BC_PRESSURE_ALARMS ) != TRUE ) + if ( getTestConfigStatus( TEST_CONFIG_DD_DISABLE_BC_PRESSURE_ALARMS ) != TRUE ) { if ( ( TRUE != isPressureDroppedDuringFill ) && ( TRUE == isFirstCycleBCSwitchingCompleted ) ) { @@ -743,7 +743,7 @@ // Check fresh and spent dialysate pressure in range if ( ( ( freshDialPressure >= FRESH_DIAL_PRESSURE_MIN_PSIG ) && ( freshDialPressure <= FRESH_DIAL_PRESSURE_MAX_PSIG ) ) && ( ( spentDialPressure >= SPENT_DIAL_PRESSURE_MIN_PSIG ) && ( spentDialPressure <= SPENT_DIAL_PRESSURE_MAX_PSIG ) ) || - ( TRUE == getBalChamberSwitchingOnlyStatus() ) || ( TRUE == getTestConfigStatus( TEST_CONFIG_DISABLE_BC_PRESSURE_ALARMS ) ) ) + ( TRUE == getBalChamberSwitchingOnlyStatus() ) || ( TRUE == getTestConfigStatus( TEST_CONFIG_DD_DISABLE_BC_PRESSURE_ALARMS ) ) ) { // Valve control for state 2 fill valveControlForBCState2FillStart(); @@ -765,7 +765,7 @@ } else { - if ( getTestConfigStatus( TEST_CONFIG_DISABLE_BC_PRESSURE_ALARMS ) != TRUE ) + if ( getTestConfigStatus( TEST_CONFIG_DD_DISABLE_BC_PRESSURE_ALARMS ) != TRUE ) { //Alarm when pressure is not in range SET_ALARM_WITH_2_F32_DATA( ALARM_ID_DD_BC_STATE2_FILL_START_PRESSURE_OUT_OF_RANGE, freshDialPressure, spentDialPressure ); @@ -790,7 +790,7 @@ spentDialPressure = getFilteredPressure( D51_PRES ); // Check fresh and spent dialysate pressure back in range to indicate fill complete. - if ( getTestConfigStatus( TEST_CONFIG_DISABLE_BC_PRESSURE_ALARMS ) != TRUE ) + if ( getTestConfigStatus( TEST_CONFIG_DD_DISABLE_BC_PRESSURE_ALARMS ) != TRUE ) { if ( ( ( freshDialPressure >= FRESH_DIAL_PRESSURE_MIN_PSIG ) && ( freshDialPressure <= FRESH_DIAL_PRESSURE_MAX_PSIG ) ) && ( ( spentDialPressure >= SPENT_DIAL_PRESSURE_MIN_PSIG ) && ( spentDialPressure <= SPENT_DIAL_PRESSURE_MAX_PSIG ) ) ) @@ -847,7 +847,7 @@ // Pressure alarm check if ( TRUE != getBalChamberSwitchingOnlyStatus() ) { - if ( getTestConfigStatus( TEST_CONFIG_DISABLE_BC_PRESSURE_ALARMS ) != TRUE ) + if ( getTestConfigStatus( TEST_CONFIG_DD_DISABLE_BC_PRESSURE_ALARMS ) != TRUE ) { if ( ( TRUE != isPressureDroppedDuringFill ) && ( TRUE == isFirstCycleBCSwitchingCompleted ) ) { Index: firmware/App/Modes/ModePreGenDialysate.c =================================================================== diff -u -re0c913b397af3122889433d855c5c5fb328f9865 -r93282b8a03682015a0c70b94d95d387204e10a89 --- firmware/App/Modes/ModePreGenDialysate.c (.../ModePreGenDialysate.c) (revision e0c913b397af3122889433d855c5c5fb328f9865) +++ firmware/App/Modes/ModePreGenDialysate.c (.../ModePreGenDialysate.c) (revision 93282b8a03682015a0c70b94d95d387204e10a89) @@ -266,7 +266,7 @@ //TODO: Change to wet self test state later DD_PRE_GEN_DIALYSATE_STATE_T state = DD_PRE_GEN_DRY_BICART_FILL_CHECK; - if ( getTestConfigStatus( TEST_CONFIG_ENABLE_DRY_BICART_FILL ) == TRUE ) + if ( getTestConfigStatus( TEST_CONFIG_DD_ENABLE_DRY_BICART_FILL ) == TRUE ) { setBicartFillRequested();