Index: firmware/App/Controllers/DryBiCart.c =================================================================== diff -u -r4e082a10d0a07d9f4b042d1c3f1e957d5ce6ec58 -rab0f35ac8202c2ebb22039c4d8d50f1f1b15556c --- firmware/App/Controllers/DryBiCart.c (.../DryBiCart.c) (revision 4e082a10d0a07d9f4b042d1c3f1e957d5ce6ec58) +++ firmware/App/Controllers/DryBiCart.c (.../DryBiCart.c) (revision ab0f35ac8202c2ebb22039c4d8d50f1f1b15556c) @@ -83,9 +83,6 @@ #define DRY_BICART_DRAIN_COND_ZERO_THRESH 0.05F ///< Zero conductivity threshold #define DRY_BICART_DRAIN_COND_STABLE_SAMPLES 10U ///< De-bounce samples (10*50ms=500ms) -// depressurize -#define DRY_BICART_DEPRESSURIZE_VENT_MAX_TIME_MS ( 10 * MS_PER_SECOND ) ///< Max vent time to depressurize the dry bicart. - /// Payload record structure for dry bicart fill request typedef struct { @@ -1529,14 +1526,6 @@ state = DRY_BICART_DEPRESSURIZE_VENT_END_STATE; } } - else if ( ( TRUE == didTimeout( dryBiCarbSypplyVentStartTime, DRY_BICART_DEPRESSURIZE_VENT_MAX_TIME_MS ) ) ) - { - setValveState( D85_VALV, VALVE_STATE_CLOSED ); - dryBiCartPersistenceStartTime = 0; - dryBiCartFillVentTimeOut = TRUE; - - state = DRY_BICART_DEPRESSURIZE_VENT_END_STATE; - } else { dryBiCartPersistenceStartTime = 0;