Index: firmware/App/Controllers/DryBiCart.c =================================================================== diff -u -r847659965e5823e2e25db62c14f4611db9527f8a -re3e3cb50025380fd3f211bc4f020f8fbdfb7f4ec --- firmware/App/Controllers/DryBiCart.c (.../DryBiCart.c) (revision 847659965e5823e2e25db62c14f4611db9527f8a) +++ firmware/App/Controllers/DryBiCart.c (.../DryBiCart.c) (revision e3e3cb50025380fd3f211bc4f020f8fbdfb7f4ec) @@ -52,7 +52,7 @@ // Dry Bicart Fill #define DRY_BICART_FILL_COMPLETE_TIME_MS ( 1 * MS_PER_SECOND ) ///< Wait time to reset the request flag after fill complete -#define DRY_BICART_WCID2_FILL_COMPLETE_PRESSURE_PSI ( 3.0F + WCDI2_PRESSURE_OFFSET_PSI ) ///< Maximum pressure reached to indicate the dry bicart fill being completed for WCID2 +#define DRY_BICART_WCID2_FILL_COMPLETE_PRESSURE_PSI ( 10.0F + WCDI2_PRESSURE_OFFSET_PSI ) ///< Maximum pressure reached to indicate the dry bicart fill being completed for WCID2 #define DRY_BICART_WCID1_FILL_COMPLETE_PRESSURE_PSI ( 10.0F + WCID1_PRESSURE_OFFSET_PSI ) ///< Maximum pressure reached to indicate the dry bicart fill being completed for WCID1. #define GET_FILL_COMPLETE_PRESSURE_PSI (getTestConfigStatus( TEST_CONFIG_DD_ENABLE_WCID_1 ) != TRUE ? \ DRY_BICART_WCID2_FILL_COMPLETE_PRESSURE_PSI : DRY_BICART_WCID1_FILL_COMPLETE_PRESSURE_PSI ) ///< Get fill complete pressure based on the WCID configuration @@ -1298,7 +1298,7 @@ { BICARB_CHAMBER_FILL_EXEC_STATE_T state = BICARB_CARTRIDGE_FILL_WATER_END_STATE; F32 d66Pressure = getFilteredPressure( D66_PRES ); - F32 fillCompletePressure = GET_FILL_COMPLETE_PRESSURE_PSI; + F32 fillCompletePressure = DRY_BICART_FILL_COMPLETE_SUPPLY_PRESSURE_PSI; LVL_STATE_T bicarbChamberLevel = getBicarbChamberLevelStatus(); if ( FALSE == isDryBicartChamberFillWindowOpen() )