Index: firmware/App/Controllers/ConcentratePumps.c =================================================================== diff -u -r0c66a3534801f2ba380e95b732e05b7cfee21772 -rd70d93528b59ef7b09032a3bf4128bee0ff6e297 --- firmware/App/Controllers/ConcentratePumps.c (.../ConcentratePumps.c) (revision 0c66a3534801f2ba380e95b732e05b7cfee21772) +++ firmware/App/Controllers/ConcentratePumps.c (.../ConcentratePumps.c) (revision d70d93528b59ef7b09032a3bf4128bee0ff6e297) @@ -8,7 +8,7 @@ * @file ConcentratePumps.c * * @author (last) Dara Navaei -* @date (last) 21-Dec-2022 +* @date (last) 18-Jan-2023 * * @author (original) Quang Nguyen * @date (original) 22-Oct-2020 @@ -210,14 +210,14 @@ #ifndef _RELEASE_ if ( getSoftwareConfigStatus( SW_CONFIG_DISABLE_CONCENTRATE_PUMPS_PARK ) != SW_CONFIG_ENABLE_VALUE ) +#endif { BOOL isAcidPumpParked = getMeasuredPumpSpeed( CONCENTRATEPUMPS_CP1_ACID ); BOOL isBicarbPumpParked = getMeasuredPumpSpeed( CONCENTRATEPUMPS_CP2_BICARB ); checkPersistentAlarm( ALARM_ID_DG_ACID_CONCENTRATE_PUMP_PARK_FAULT, isAcidPumpParked, isAcidPumpParked, CONENTREATE_PUMP_PARK_FAULT_TIMEOUT_MS ); checkPersistentAlarm( ALARM_ID_DG_BICARB_CONCENTRATE_PUMP_PARK_FAULT, isBicarbPumpParked, isBicarbPumpParked, CONENTREATE_PUMP_PARK_FAULT_TIMEOUT_MS ); } -#endif if ( ++concentratePumpMonitorTimerCounter >= getU32OverrideValue( &concentratePumpDataPublishInterval ) ) {