Index: firmware/App/Controllers/ConcentratePumps.c =================================================================== diff -u -r5d82009d1baa2b52122065934481745bf4de223b -ra1877093454dc5252c43f5ab11b005a85e293167 --- firmware/App/Controllers/ConcentratePumps.c (.../ConcentratePumps.c) (revision 5d82009d1baa2b52122065934481745bf4de223b) +++ firmware/App/Controllers/ConcentratePumps.c (.../ConcentratePumps.c) (revision a1877093454dc5252c43f5ab11b005a85e293167) @@ -37,7 +37,6 @@ #define CONCENTRATE_PUMP_SPEED_INCREMENT 8.0 ///< Speed increase (mL/min) when controlling concentrate pump to target step speed. #define CONCENTRATE_PUMP_MIN_SPEED 3.0 ///< Minimum speed for concentrate pump in mL per min. -#define CONCENTRATE_PUMP_MAX_SPEED 49.0 ///< Maximum speed for concentrate pump in mL per min. #define CONCENTRATE_PUMP_ERROR_TOLERANCE 0.02 ///< Measured speed needs to be within 2% of commanded speed. #define CONCENTRATE_PUMP_ZERO_FLOW_RATE 0xFFFF ///< Pulse width value when zero flow rate or pump is off.