Index: firmware/App/Controllers/ConcentratePumps.c =================================================================== diff -u -r6499ea25921fcf67826fa0c35bb03caf411ba542 -r4d364420c08743808916d42276823773d9329682 --- firmware/App/Controllers/ConcentratePumps.c (.../ConcentratePumps.c) (revision 6499ea25921fcf67826fa0c35bb03caf411ba542) +++ firmware/App/Controllers/ConcentratePumps.c (.../ConcentratePumps.c) (revision 4d364420c08743808916d42276823773d9329682) @@ -209,14 +209,14 @@ #ifndef _RELEASE_ if ( getSoftwareConfigStatus( SW_CONFIG_DISABLE_CONCENTRATE_PUMPS_PARK ) != SW_CONFIG_ENABLE_VALUE ) +#endif { BOOL isAcidPumpParked = getMeasuredPumpSpeed( CONCENTRATEPUMPS_CP1_ACID ); BOOL isBicarbPumpParked = getMeasuredPumpSpeed( CONCENTRATEPUMPS_CP2_BICARB ); checkPersistentAlarm( ALARM_ID_DG_ACID_CONCENTRATE_PUMP_PARK_FAULT, isAcidPumpParked, isAcidPumpParked, CONENTREATE_PUMP_PARK_FAULT_TIMEOUT_MS ); checkPersistentAlarm( ALARM_ID_DG_BICARB_CONCENTRATE_PUMP_PARK_FAULT, isBicarbPumpParked, isBicarbPumpParked, CONENTREATE_PUMP_PARK_FAULT_TIMEOUT_MS ); } -#endif if ( ++concentratePumpMonitorTimerCounter >= getU32OverrideValue( &concentratePumpDataPublishInterval ) ) { @@ -562,7 +562,7 @@ // Park concentrate pump too if requested #ifndef _RELEASE_ - if ( getSoftwareConfigStatus( SW_CONFIG_DISABLE_CONCENTRATE_PUMPS_PARK ) != SW_CONFIG_ENABLE_VALUE ) + if ( SW_CONFIG_DISABLE_VALUE == getSoftwareConfigStatus( SW_CONFIG_DISABLE_CONCENTRATE_PUMPS_PARK ) ) { if ( HW_CONFIG_BETA == getHardwareConfigStatus() ) {