Index: firmware/App/Controllers/ConcentratePumps.c =================================================================== diff -u -r6499ea25921fcf67826fa0c35bb03caf411ba542 -re107dacab9f6c78574f1c26e1004e0dbdc8dba03 --- firmware/App/Controllers/ConcentratePumps.c (.../ConcentratePumps.c) (revision 6499ea25921fcf67826fa0c35bb03caf411ba542) +++ firmware/App/Controllers/ConcentratePumps.c (.../ConcentratePumps.c) (revision e107dacab9f6c78574f1c26e1004e0dbdc8dba03) @@ -209,14 +209,14 @@ #ifndef _RELEASE_ if ( getSoftwareConfigStatus( SW_CONFIG_DISABLE_CONCENTRATE_PUMPS_PARK ) != SW_CONFIG_ENABLE_VALUE ) +#endif { BOOL isAcidPumpParked = getMeasuredPumpSpeed( CONCENTRATEPUMPS_CP1_ACID ); BOOL isBicarbPumpParked = getMeasuredPumpSpeed( CONCENTRATEPUMPS_CP2_BICARB ); checkPersistentAlarm( ALARM_ID_DG_ACID_CONCENTRATE_PUMP_PARK_FAULT, isAcidPumpParked, isAcidPumpParked, CONENTREATE_PUMP_PARK_FAULT_TIMEOUT_MS ); checkPersistentAlarm( ALARM_ID_DG_BICARB_CONCENTRATE_PUMP_PARK_FAULT, isBicarbPumpParked, isBicarbPumpParked, CONENTREATE_PUMP_PARK_FAULT_TIMEOUT_MS ); } -#endif if ( ++concentratePumpMonitorTimerCounter >= getU32OverrideValue( &concentratePumpDataPublishInterval ) ) {