Index: firmware/App/Controllers/ConcentratePumps.c =================================================================== diff -u -r6743b1ba8056267f3a1669a3773c9f9995c0a788 -r8ae4fbcd07215613d418e63a54d2a6fa80a83a4a --- firmware/App/Controllers/ConcentratePumps.c (.../ConcentratePumps.c) (revision 6743b1ba8056267f3a1669a3773c9f9995c0a788) +++ firmware/App/Controllers/ConcentratePumps.c (.../ConcentratePumps.c) (revision 8ae4fbcd07215613d418e63a54d2a6fa80a83a4a) @@ -203,6 +203,9 @@ NUMBER_OF_ACID_AND_BICARB_NV_DATA_TO_CHECK, ALARM_ID_DG_ACID_CONCENTRATE_INVALID_CAL_RECORD ); getNVRecord2Driver( GET_CAL_BICARB_CONCENTRATES, (U08*)&bicarbConcentrateCalRecord, sizeof( bicarbConcentrateCalRecord ), NUMBER_OF_ACID_AND_BICARB_NV_DATA_TO_CHECK, ALARM_ID_DG_BICARB_CONCENTRATE_INVALID_CAL_RECORD ); + + getNVRecord2Driver( GET_CAL_CONCENTRATE_PUMPS_RECORD, (U08*)&concentratePumpsCalRecord, sizeof( DG_CONC_PUMPS_CAL_RECORD_T ), + NUM_OF_CAL_DATA_DG_CONC_PUMPS, ALARM_ID_DG_CONCENTRATE_PUMPS_INVALID_CAL_RECORD ); } // Calculate pump speed for each defined pump @@ -351,7 +354,7 @@ NUMBER_OF_ACID_AND_BICARB_NV_DATA_TO_CHECK, ALARM_ID_DG_BICARB_CONCENTRATE_INVALID_CAL_RECORD ); calStatus |= getNVRecord2Driver( GET_CAL_CONCENTRATE_PUMPS_RECORD, (U08*)&concentratePumpsCalRecord, sizeof( DG_CONC_PUMPS_CAL_RECORD_T ), - NUM_OF_CAL_DATA_DG_CONC_PUMPS, ALARM_ID_NO_ALARM ); + NUM_OF_CAL_DATA_DG_CONC_PUMPS, ALARM_ID_DG_CONCENTRATE_PUMPS_INVALID_CAL_RECORD ); result = ( TRUE == calStatus ? SELF_TEST_STATUS_PASSED : SELF_TEST_STATUS_FAILED );