Index: firmware/App/Controllers/ConcentratePumps.c =================================================================== diff -u -r77e697d1d8fc4770c6eab15794da7e96f93b847a -rc7e3b16c5dfe2f77e8853262c1452734f80f9f2c --- firmware/App/Controllers/ConcentratePumps.c (.../ConcentratePumps.c) (revision 77e697d1d8fc4770c6eab15794da7e96f93b847a) +++ firmware/App/Controllers/ConcentratePumps.c (.../ConcentratePumps.c) (revision c7e3b16c5dfe2f77e8853262c1452734f80f9f2c) @@ -217,8 +217,8 @@ if ( getSoftwareConfigStatus( SW_CONFIG_DISABLE_CONCENTRATE_PUMPS_PARK ) != SW_CONFIG_ENABLE_VALUE ) #endif { - BOOL isAcidPumpParked = (U32)getFPGAAcidPumpParkFault(); - BOOL isBicarbPumpParked = (U32)getFPGABicarbPumpParkFault(); + BOOL isAcidPumpParked = getConcPumpParkIsFaulted( CONCENTRATEPUMPS_CP1_ACID ); + BOOL isBicarbPumpParked = getConcPumpParkIsFaulted( CONCENTRATEPUMPS_CP2_BICARB ); checkPersistentAlarm( ALARM_ID_DG_ACID_CONCENTRATE_PUMP_PARK_FAULT, isAcidPumpParked, isAcidPumpParked, CONENTREATE_PUMP_PARK_FAULT_TIMEOUT_MS ); checkPersistentAlarm( ALARM_ID_DG_BICARB_CONCENTRATE_PUMP_PARK_FAULT, isBicarbPumpParked, isBicarbPumpParked, CONENTREATE_PUMP_PARK_FAULT_TIMEOUT_MS );