Index: firmware/App/Controllers/ConcentratePumps.c =================================================================== diff -u -r986abcfcf047822cad1a10c1ee0924a80dd5f512 -r7d200a6cc3958f49c9a84cc5372a04228876d2fe --- firmware/App/Controllers/ConcentratePumps.c (.../ConcentratePumps.c) (revision 986abcfcf047822cad1a10c1ee0924a80dd5f512) +++ firmware/App/Controllers/ConcentratePumps.c (.../ConcentratePumps.c) (revision 7d200a6cc3958f49c9a84cc5372a04228876d2fe) @@ -543,7 +543,7 @@ *************************************************************************/ static void calcMeasuredPumpsSpeed( CONCENTRATE_PUMPS_T pumpId, U16 pulseWidthCount ) { - F32 pulseWidthInSecond = (F32)( pulseWidthCount * CONCENTRATE_PUMP_HALL_SENSE_PERIOD_RESOLUTION ) / US_PER_SECOND; + F32 pulseWidthInSecond = ( (F32)pulseWidthCount * CONCENTRATE_PUMP_HALL_SENSE_PERIOD_RESOLUTION ) / US_PER_SECOND; BOOL isPulseWidthOut = ( pulseWidthInSecond <= CONCENTRATE_PUMP_MIN_ALLOWED_HALL_SENSOR_COUNT ? TRUE : FALSE ); //checkPersistentAlarm( ALARM_ID_DG_CONC_PUMP_HALL_SENSOR_OUT_OF_RANGE, isPulseWidthOut, pulseWidthInSecond, CONCENTRATE_PUMP_MIN_ALLOWED_HALL_SENSOR_COUNT );