Index: firmware/App/Controllers/ConcentratePumps.c =================================================================== diff -u -ra3916397312e1dc73efe99d6b3569b7387a9419b -r77e697d1d8fc4770c6eab15794da7e96f93b847a --- firmware/App/Controllers/ConcentratePumps.c (.../ConcentratePumps.c) (revision a3916397312e1dc73efe99d6b3569b7387a9419b) +++ firmware/App/Controllers/ConcentratePumps.c (.../ConcentratePumps.c) (revision 77e697d1d8fc4770c6eab15794da7e96f93b847a) @@ -217,8 +217,8 @@ if ( getSoftwareConfigStatus( SW_CONFIG_DISABLE_CONCENTRATE_PUMPS_PARK ) != SW_CONFIG_ENABLE_VALUE ) #endif { - BOOL isAcidPumpParked = getMeasuredPumpSpeed( CONCENTRATEPUMPS_CP1_ACID ); - BOOL isBicarbPumpParked = getMeasuredPumpSpeed( CONCENTRATEPUMPS_CP2_BICARB ); + BOOL isAcidPumpParked = (U32)getFPGAAcidPumpParkFault(); + BOOL isBicarbPumpParked = (U32)getFPGABicarbPumpParkFault(); checkPersistentAlarm( ALARM_ID_DG_ACID_CONCENTRATE_PUMP_PARK_FAULT, isAcidPumpParked, isAcidPumpParked, CONENTREATE_PUMP_PARK_FAULT_TIMEOUT_MS ); checkPersistentAlarm( ALARM_ID_DG_BICARB_CONCENTRATE_PUMP_PARK_FAULT, isBicarbPumpParked, isBicarbPumpParked, CONENTREATE_PUMP_PARK_FAULT_TIMEOUT_MS );