Index: firmware/App/Controllers/ConductivitySensors.c =================================================================== diff -u -r54bdfb2c6f746097467152eabf62961b0a72542f -raa167d853a75c5ba1627f3ff51fd9a25542ab82b --- firmware/App/Controllers/ConductivitySensors.c (.../ConductivitySensors.c) (revision 54bdfb2c6f746097467152eabf62961b0a72542f) +++ firmware/App/Controllers/ConductivitySensors.c (.../ConductivitySensors.c) (revision aa167d853a75c5ba1627f3ff51fd9a25542ab82b) @@ -51,7 +51,7 @@ #define MAX_COND_SENSOR_CPI_WARNING_HIGH_US_PER_CM 2000.0F ///< Maximum allowed high conductivity value in uS/cm. #define MIN_COND_SENSOR_CPI_WARNING_HIGH_US_PER_CM 1990.0F ///< Minimum allowed high conductivity value in uS/cm. -#define MIN_CPI_INLET_ALARM_RECOVERY_OFFSET_US_PER_CM 20.0F ///< Minimum inlet water conductivity recovery offset value in uS/cm. +#define MIN_CPI_INLET_ALARM_RECOVERY_OFFSET_US_PER_CM 10.0F ///< Minimum inlet water conductivity recovery offset value in uS/cm. #define MAX_RO_ONLY_COND_SENSOR_CPI_HIGH_US_PER_CM 100.0F ///< Maximum RO only mode high conductivity value in uS/cm. #define MIN_RO_ONLY_COND_SENSOR_CPI_HIGH_US_PER_CM 90.0F ///< Minimum RO only mode high conductivity value in uS/cm.