Index: firmware/App/Controllers/ConductivitySensors.c =================================================================== diff -u -r6085153faf2a176c51a34653bdd3a36a4699e528 -r86617288c49b57eca42cf2d0b67ade76e69f96bd --- firmware/App/Controllers/ConductivitySensors.c (.../ConductivitySensors.c) (revision 6085153faf2a176c51a34653bdd3a36a4699e528) +++ firmware/App/Controllers/ConductivitySensors.c (.../ConductivitySensors.c) (revision 86617288c49b57eca42cf2d0b67ade76e69f96bd) @@ -44,17 +44,6 @@ #define COND_SENSOR_CPI_MAX_VALUE 2000 ///< Maximum inlet water conductivity. #define COND_SENSOR_CPI_MIN_VALUE 100 ///< Minimum inlet water conductivity. -#define MIN_ACID_CONCENTRATE_CONDUCTIVITY 11000 ///< Minimum conductivity for acid concentrate. -#define MAX_ACID_CONCENTRATE_CONDUCTIVITY 11850 ///< Maximum conductivity for acid concentrate. - -#define MIN_BICARB_CONCENTRATE_CONDUCTIVITY 2000 ///< Minimum conductivity for bicarbonate concentrate. -#define MAX_BICARB_CONCENTRATE_CONDUCTIVITY 3000 ///< Maximum conductivity for bicarbonate concentrate. - -///< Minimum conductivity value for dialysate solution. -#define MIN_DIALYSATE_CONDUCTIVITY ( MIN_ACID_CONCENTRATE_CONDUCTIVITY + MIN_BICARB_CONCENTRATE_CONDUCTIVITY ) -///< Maximum conductivity value for dialysate solution. -#define MAX_DIALYSATE_CONDUCTIVITY ( MAX_ACID_CONCENTRATE_CONDUCTIVITY + MAX_BICARB_CONCENTRATE_CONDUCTIVITY ) - #define MAX_ALLOWED_UNCHANGED_CONDUCTIVITY_READS ( MS_PER_SECOND / TASK_PRIORITY_INTERVAL ) ///< New reading every 800 ms, expect to get valid new reading in 1s. #define MAX_CONDUCTIVITY_SENSOR_FAILURES 5 ///< maximum number of conductivity sensor errors within window period before alarm. #define MAX_CONDUCTIVITY_SENSOR_FAILURE_WINDOW_MS ( 60 * MS_PER_SECOND ) ///< Conductivity sensor error window.