Index: firmware/App/Controllers/ConductivitySensors.h =================================================================== diff -u -r3414e47b2664c56ab6b2f0cd5cd6b30ae9a5e333 -r66cc826068b54bc436cde0fae70a05ba6c1ac974 --- firmware/App/Controllers/ConductivitySensors.h (.../ConductivitySensors.h) (revision 3414e47b2664c56ab6b2f0cd5cd6b30ae9a5e333) +++ firmware/App/Controllers/ConductivitySensors.h (.../ConductivitySensors.h) (revision 66cc826068b54bc436cde0fae70a05ba6c1ac974) @@ -7,8 +7,8 @@ * * @file ConductivitySensors.h * -* @author (last) Hung Nguyen -* @date (last) 09-Feb-2022 +* @author (last) Dara Navaei +* @date (last) 31-Mar-2022 * * @author (original) Quang Nguyen * @date (original) 13-Jul-2020 @@ -46,17 +46,9 @@ // ********** public definitions ********** -#define MIN_ACID_CONCENTRATE_CONDUCTIVITY 11000 ///< Minimum conductivity for acid concentrate. -#define MAX_ACID_CONCENTRATE_CONDUCTIVITY 11850 ///< Maximum conductivity for acid concentrate. +#define ACID_NORMAL_CONDUCTIVITY 11645.05 ///< Acid normal operation theoretical conductivity. +#define BICARB_NORMAL_CONDUCTIVITY 13734.88 ///< Bicarb normal operation theoretical conductivity. -#define MIN_BICARB_CONCENTRATE_CONDUCTIVITY 2000 ///< Minimum conductivity for bicarbonate concentrate. -#define MAX_BICARB_CONCENTRATE_CONDUCTIVITY 3000 ///< Maximum conductivity for bicarbonate concentrate. - -///< Minimum conductivity value for dialysate solution. -#define MIN_DIALYSATE_CONDUCTIVITY ( MIN_ACID_CONCENTRATE_CONDUCTIVITY + MIN_BICARB_CONCENTRATE_CONDUCTIVITY ) -///< Maximum conductivity value for dialysate solution. -#define MAX_DIALYSATE_CONDUCTIVITY ( MAX_ACID_CONCENTRATE_CONDUCTIVITY + MAX_BICARB_CONCENTRATE_CONDUCTIVITY ) - /// Enumeration of conductivity sensors. typedef enum ConductivitySensors {