Index: firmware/App/Controllers/LoadCell.c =================================================================== diff -u -rb64c49fdcf2b6d95e61e63f8e258c4e600935bbd -rf2d00ceb224ce5f7a239379fff23ee63d919cfb7 --- firmware/App/Controllers/LoadCell.c (.../LoadCell.c) (revision b64c49fdcf2b6d95e61e63f8e258c4e600935bbd) +++ firmware/App/Controllers/LoadCell.c (.../LoadCell.c) (revision f2d00ceb224ce5f7a239379fff23ee63d919cfb7) @@ -93,3 +93,17 @@ return Load_cell_b2_ave; } +/************************************************************************* + * @brief testSetLoadCellA1Override and testResetLoadCellA1Override + * The testSetLoadCellA1Override function overrides the measured \n + * load cell A1. \n + * The testResetLoadCellA1Override function resets the override of the \n + * load cell A1. + * @details + * Inputs : none + * Outputs : raw adc value for load cell A1 + * @param value : override measured blood pump current (in mA) + * @return TRUE if override successful, FALSE if not + *************************************************************************/ +DATA_OVERRIDE_FUNC( F32, testSetMeasuredBloodPumpMCCurrentOverride, testResetMeasuredBloodPumpMCCurrentOverride, adcBloodPumpMCCurrentmA ) +