Index: firmware/App/Controllers/Pressures.c =================================================================== diff -u -r22176ce95e49213c48454f34ddf5d29b8109f2cb -r224b86c0847207c168d38fa1713a1e9116642267 --- firmware/App/Controllers/Pressures.c (.../Pressures.c) (revision 22176ce95e49213c48454f34ddf5d29b8109f2cb) +++ firmware/App/Controllers/Pressures.c (.../Pressures.c) (revision 224b86c0847207c168d38fa1713a1e9116642267) @@ -135,7 +135,6 @@ *************************************************************************/ void checkInletWaterPressure( void ) { -#ifndef DISABLE_WATER_QUALITY_CHECK #ifndef _RELEASE_ if ( getSoftwareConfigStatus( SW_CONFIG_DISABLE_WATER_QUALITY_CHECK ) != SW_CONFIG_ENABLE_VALUE ) #endif @@ -153,7 +152,6 @@ checkPersistentAlarm( ALARM_ID_INLET_WATER_PRESSURE_IN_LOW_RANGE, FALSE, pressure, MAX_INLET_WATER_PRESSURE_WARNING_LOW ); } } -#endif } /*********************************************************************//** @@ -225,7 +223,7 @@ { case PRESSURE_SELF_TEST_STATE_START: getNVRecord2Driver( GET_CAL_PRESSURE_SENOSRS, (U08*)&pressuresCalRecord, sizeof( DG_PRES_SENSORS_CAL_RECORD_T ), - NUM_OF_CAL_DATA_PRES_SENSORS, ALARM_ID_DG_PRESSURE_SENSORS_INVALID_CAL_RECORD ); + NUM_OF_CAL_DATA_PRES_SENSORS, ALARM_ID_NO_ALARM ); pressuresSelfTestState = PRESSURE_TEST_STATE_IN_PROGRESS; pressuresSelfTestResult = SELF_TEST_STATUS_IN_PROGRESS; break; @@ -291,7 +289,7 @@ if ( TRUE == isNewCalibrationRecordAvailable() ) { getNVRecord2Driver( GET_CAL_PRESSURE_SENOSRS, (U08*)&pressuresCalRecord, sizeof( DG_PRES_SENSORS_CAL_RECORD_T ), - NUM_OF_CAL_DATA_PRES_SENSORS, ALARM_ID_DG_PRESSURE_SENSORS_INVALID_CAL_RECORD ); + NUM_OF_CAL_DATA_PRES_SENSORS, ALARM_ID_NO_ALARM ); } // filter every 200ms