Index: firmware/App/Controllers/ROPump.c =================================================================== diff -u -r1a7b3fdc8c9b47ae713a7ec37670a96df7d73818 -r3f7d30b23906496854054949d4491f3bae6ef3c4 --- firmware/App/Controllers/ROPump.c (.../ROPump.c) (revision 1a7b3fdc8c9b47ae713a7ec37670a96df7d73818) +++ firmware/App/Controllers/ROPump.c (.../ROPump.c) (revision 3f7d30b23906496854054949d4491f3bae6ef3c4) @@ -179,9 +179,6 @@ initPersistentAlarm( ALARM_ID_RO_PUMP_PRESSURE_OUT_OF_RANGE, MAX_PRESSURE_OUT_OF_RANGE_PERSISTENT_INTERVAL, MAX_PRESSURE_OUT_OF_RANGE_PERSISTENT_INTERVAL ); - // Initialize the persistent alarm for ramp up to target flow timeout - initPersistentAlarm( ALARM_ID_RO_PUMP_RAMP_UP_TO_FLOW_TIMEOUT, MAX_ALLOWED_RAMP_UP_TIME, MAX_ALLOWED_RAMP_UP_TIME ); - // Initialize the persistent alarm for not turning off the pump initPersistentAlarm( ALARM_ID_RO_PUMP_OFF_FAULT, SAFETY_SHUTDOWN_TIMEOUT, SAFETY_SHUTDOWN_TIMEOUT );