Index: firmware/App/Modes/ModeChemicalDisinfect.c =================================================================== diff -u -r22176ce95e49213c48454f34ddf5d29b8109f2cb -rf91baac9704a90ab0fea1dd05e11833ffc5a5a4e --- firmware/App/Modes/ModeChemicalDisinfect.c (.../ModeChemicalDisinfect.c) (revision 22176ce95e49213c48454f34ddf5d29b8109f2cb) +++ firmware/App/Modes/ModeChemicalDisinfect.c (.../ModeChemicalDisinfect.c) (revision f91baac9704a90ab0fea1dd05e11833ffc5a5a4e) @@ -80,7 +80,7 @@ // R1 to R2 & R2 to R1 chemical disinfect circulation #define CHEM_DISINFECT_TARGET_RO_FLOW_LPM 0.8F ///< Chemical disinfect target RO flow rate in L/min. #define CHEM_DISINFECT_MAX_RO_PRESSURE_PSI 130 ///< Chemical disinfect maximum RO pressure in psi. -#define CHEM_DISINFECT_TARGET_DRAIN_PRES_PSI 10.0 ///< Chemical disinfect target drain outlet pressure in psi. +#define CHEM_DISINFECT_TARGET_DRAIN_PRES_PSI 12.0 ///< Chemical disinfect target drain outlet pressure in psi. #define CHEM_DISINFECT_TIME_MS ( 36 * SEC_PER_MIN * MS_PER_SECOND ) ///< Chemical disinfect time for each section in milliseconds. TODO original time was 36 minutes #define CHEM_DISINFECT_START_TEMP_TIMOUT_MS ( 4 * MIN_PER_HOUR * SEC_PER_MIN * MS_PER_SECOND ) ///< Chemical disinfect reaching to minimum temperature timeout in milliseconds. #define RSRVRS_TARGET_VOL_OUT_TIMEOUT_MS ( 0.5F * SEC_PER_MIN * MS_PER_SECOND ) ///< Reservoirs 1 & 2 maximum volume out of range timeout during chemical disinfect. TODO change this to 5 seconds