Index: firmware/App/Modes/ModeChemicalDisinfect.c =================================================================== diff -u -ra727129f5f547fc7a46f481417f7b926454d3e97 -r0803f828b81e046166457564650acf6a9bbd3cc6 --- firmware/App/Modes/ModeChemicalDisinfect.c (.../ModeChemicalDisinfect.c) (revision a727129f5f547fc7a46f481417f7b926454d3e97) +++ firmware/App/Modes/ModeChemicalDisinfect.c (.../ModeChemicalDisinfect.c) (revision 0803f828b81e046166457564650acf6a9bbd3cc6) @@ -999,12 +999,11 @@ static DG_CHEM_DISINFECT_STATE_T handleChemicalDisinfectPrimeAcidLineState( void ) { DG_CHEM_DISINFECT_STATE_T state = DG_CHEM_DISINFECT_STATE_PRIME_ACID_LINE; + F32 cd2Conductivity = getConductivityValue( (U32)CONDUCTIVITYSENSORS_CD2_SENSOR ); // Set the chemical disinfect that is published on the UI chemDisinfectUIState = CHEM_DISINFECT_UI_STATE_MIX_WATER_AND_ACID; - F32 cd2Conductivity = getConductivityValue( (U32)CONDUCTIVITYSENSORS_CD2_SENSOR ); - if ( cd2Conductivity <= MIN_ACID_CONDUCTIVITY_US_PER_CM ) { primeAcidSteadyStateCounter = 0;