Index: firmware/App/Modes/ModeChemicalDisinfect.c =================================================================== diff -u -re258c4013bdffbf4f9e77ab9ef35fd67e3d1bb8f -reeaab2a9a685fb18845888d91040d3aa01977f6d --- firmware/App/Modes/ModeChemicalDisinfect.c (.../ModeChemicalDisinfect.c) (revision e258c4013bdffbf4f9e77ab9ef35fd67e3d1bb8f) +++ firmware/App/Modes/ModeChemicalDisinfect.c (.../ModeChemicalDisinfect.c) (revision eeaab2a9a685fb18845888d91040d3aa01977f6d) @@ -7,8 +7,8 @@ * * @file ModeChemicalDisinfect.c * -* @author (last) Dara Navaei -* @date (last) 30-Mar-2023 +* @author (last) Michael Garthwaite +* @date (last) 16-May-2023 * * @author (original) Sean * @date (original) 04-Apr-2020 @@ -293,9 +293,12 @@ deenergizeActuators( NO_PARK_CONC_PUMPS ); initChemicalDisinfectMode(); - + setCurrentSubState( NO_SUB_STATE ); setCPLDCleanLEDColor( CPLD_CLEAN_LED_YELLOW ); + // Set CD1 and CD2 conductivity sensors calibration table to be picked from the chemical disinfect calibration record + setCondcutivitySensorCalTable( CONDUCTIVITYSENSORS_CD2_SENSOR, CAL_DATA_CD2_COND_SENSOR_CHEM_DISINFECT ); + #ifndef _RELEASE_ if ( nelsonSupport != NELSON_CHEM_DISINFECT ) #endif