Index: firmware/App/Modes/ModeChemicalDisinfectFlush.c =================================================================== diff -u -rbec5bb7634100c96341d19cf89e540a01aa86b1d -rc940dea020a10f2bab23ae92aa6932c375c26dd4 --- firmware/App/Modes/ModeChemicalDisinfectFlush.c (.../ModeChemicalDisinfectFlush.c) (revision bec5bb7634100c96341d19cf89e540a01aa86b1d) +++ firmware/App/Modes/ModeChemicalDisinfectFlush.c (.../ModeChemicalDisinfectFlush.c) (revision c940dea020a10f2bab23ae92aa6932c375c26dd4) @@ -8,7 +8,7 @@ * @file ModeChemicalDisinfectFlush.c * * @author (last) Dara Navaei -* @date (last) 04-May-2023 +* @date (last) 11-May-2023 * * @author (original) Dara Navaei * @date (original) 15-Nov-2022 @@ -200,6 +200,11 @@ setCPLDCleanLEDColor( CPLD_CLEAN_LED_YELLOW ); + // Upon transition to mode chemical flush set CD1 and CD2 calibration records to be picked to the normal + // table. If the chemical disinfect is successful, mode chemical disinfect flush is automatically started + setCondcutivitySensorCalTable( CONDUCTIVITYSENSORS_CD1_SENSOR, CAL_DATA_CD1_COND_SENSOR ); + setCondcutivitySensorCalTable( CONDUCTIVITYSENSORS_CD2_SENSOR, CAL_DATA_CD2_COND_SENSOR ); + activateAlarmNoData( ALARM_ID_DG_CHEM_DISINFECT_FLUSH_REMOVE_ACID ); return chemDisinfectFlushState;