Index: firmware/App/Modes/ModeChemicalDisinfectFlush.c =================================================================== diff -u -rd19a072db6880577f677093929b53445e34a33e3 -r8c5fef6b7ab382db81084a3abe9604e01882aef9 --- firmware/App/Modes/ModeChemicalDisinfectFlush.c (.../ModeChemicalDisinfectFlush.c) (revision d19a072db6880577f677093929b53445e34a33e3) +++ firmware/App/Modes/ModeChemicalDisinfectFlush.c (.../ModeChemicalDisinfectFlush.c) (revision 8c5fef6b7ab382db81084a3abe9604e01882aef9) @@ -8,7 +8,7 @@ * @file ModeChemicalDisinfectFlush.c * * @author (last) Dara Navaei -* @date (last) 04-May-2023 +* @date (last) 11-May-2023 * * @author (original) Dara Navaei * @date (original) 15-Nov-2022 @@ -200,6 +200,11 @@ setCurrentSubState( NO_SUB_STATE ); setCPLDCleanLEDColor( CPLD_CLEAN_LED_YELLOW ); + // Upon transition to mode chemical flush set CD1 and CD2 calibration records to be picked to the normal + // table. If the chemical disinfect is successful, mode chemical disinfect flush is automatically started + setCondcutivitySensorCalTable( CONDUCTIVITYSENSORS_CD1_SENSOR, CAL_DATA_CD1_COND_SENSOR ); + setCondcutivitySensorCalTable( CONDUCTIVITYSENSORS_CD2_SENSOR, CAL_DATA_CD2_COND_SENSOR ); + activateAlarmNoData( ALARM_ID_DG_CHEM_DISINFECT_FLUSH_REMOVE_ACID ); return chemDisinfectFlushState;