Index: firmware/App/Modes/ModeDrain.c =================================================================== diff -u -ra9315539f527b92523b1598ff91e47db4d71dae2 -rc7ed4f90b600fe684097654f150e4972646ffec9 --- firmware/App/Modes/ModeDrain.c (.../ModeDrain.c) (revision a9315539f527b92523b1598ff91e47db4d71dae2) +++ firmware/App/Modes/ModeDrain.c (.../ModeDrain.c) (revision c7ed4f90b600fe684097654f150e4972646ffec9) @@ -19,6 +19,7 @@ #include "ConductivitySensors.h" #include "DrainPump.h" #include "Heaters.h" +#include "ModeFill.h" #include "ModeDrain.h" #include "OperationModes.h" #include "Pressures.h" @@ -242,6 +243,19 @@ setValveState( VRD2, VALVE_STATE_CLOSED ); } + // check for empty bottles at the end of drain + if ( TRUE == isEmptyAcidBottle() ) // is acid volume < 10% + { + setThisFisrtFillFlag( TRUE ); // indicates bottles need prime + activateAlarmNoData( ALARM_ID_DG_ACID_BOTTLE_LOW_VOLUME ); // trigger empty acid bottle alarm + } + + if ( TRUE == isEmptyBicarbBottle() ) + { + setThisFisrtFillFlag( TRUE ); + activateAlarmNoData( ALARM_ID_DG_BICARB_BOTTLE_LOW_VOLUME ); + } + if ( TRUE == isReservoirTarePending() ) { // Tare reservoir load cells at empty if requested result = DG_DRAIN_STATE_TARE; @@ -251,9 +265,7 @@ requestNewOperationMode( DG_MODE_GENE ); } } - - // Drain timed out raise the alarm - if ( TRUE == didTimeout( dialysateDrainStartTime, DIALYSATE_DRAIN_TIME_OUT ) ) + else if ( TRUE == didTimeout( dialysateDrainStartTime, DIALYSATE_DRAIN_TIME_OUT ) ) // Drain timed out raise the alarm { activateAlarmNoData( ALARM_ID_DG_DIALYSATE_DRAIN_TIME_OUT ); }