Index: firmware/App/Modes/ModeFill.c =================================================================== diff -u -r1722e975682c9e41e0b75551ada1139d3dc522a6 -ra0966852dd6f9f8e70289b7ea7588677967f3f2f --- firmware/App/Modes/ModeFill.c (.../ModeFill.c) (revision 1722e975682c9e41e0b75551ada1139d3dc522a6) +++ firmware/App/Modes/ModeFill.c (.../ModeFill.c) (revision a0966852dd6f9f8e70289b7ea7588677967f3f2f) @@ -564,23 +564,23 @@ DG_FILL_MODE_STATE_T result = DG_FILL_MODE_STATE_TEST_BICARB_CONDUCTIVITY; F32 bicarbConductivity = getConductivityValue( CONDUCTIVITYSENSORS_CD2_SENSOR ); -#ifndef DISABLE_BICARB_CONDUCTIVITY_TEST totalBicarbConductivity += bicarbConductivity; bicarbConductivitySampleCount++; if ( TRUE == didTimeout( concentrateTestStartTime, CONCENTRATE_TEST_TIME_OUT_MS ) ) { averageBicarbConductivity = totalBicarbConductivity / bicarbConductivitySampleCount; +#ifndef DISABLE_BICARB_CONDUCTIVITY_TEST if ( TRUE == isValueWithinPercentRange( averageBicarbConductivity, BICARB_TEST_CD2_TCD, FIVE_PERCENT_FACTOR ) ) +#endif { // Initialization requestConcentratePumpOff( CONCENTRATEPUMPS_CP2_BICARB ); totalBicarbConductivity = 0.0; totalAcidConductivity = 0.0; bicarbConductivitySampleCount = 0; acidConductivitySampleCount = 0; -#endif // Set pumps flow rate to prepare for acid conductivity testing setROPumpTargetFlowRateLPM( RO_PUMP_800_ML_PER_MIN / MILLILITERS_PER_LITER, TARGET_RO_PRESSURE_PSI ); @@ -590,15 +590,15 @@ concentrateTestStartTime = getMSTimerCount(); result = DG_FILL_MODE_STATE_TEST_ACID_CONDUCTIVITY; -#ifndef DISABLE_BICARB_CONDUCTIVITY_TEST } +#ifndef DISABLE_BICARB_CONDUCTIVITY_TEST else { // Bicarb test failed, go to fault SET_ALARM_WITH_1_F32_DATA( ALARM_ID_BICARB_CONDUCTIVITY_OUT_OF_RANGE, averageBicarbConductivity ); } - } #endif + } return result; } @@ -619,7 +619,6 @@ F32 acidConductivity = getConductivityValue( CONDUCTIVITYSENSORS_CD1_SENSOR ); F32 bicarbConductivity = getConductivityValue( CONDUCTIVITYSENSORS_CD2_SENSOR ); -#ifndef DISABLE_ACID_CONDUCTIVITY_TEST totalBicarbConductivity += bicarbConductivity; bicarbConductivitySampleCount++; @@ -632,11 +631,11 @@ averageAcidConductivity = totalAcidConductivity / acidConductivitySampleCount; pctDiffInConductivity = fabs( 2.0 * ( averageAcidConductivity - averageBicarbConductivity ) / ( averageAcidConductivity + averageBicarbConductivity ) ); - +#ifndef DISABLE_ACID_CONDUCTIVITY_TEST if ( ( TRUE == isValueWithinPercentRange( averageAcidConductivity, ACID_TEST_CD2_TCD, FIVE_PERCENT_FACTOR ) ) && ( pctDiffInConductivity < FIVE_PERCENT_FACTOR ) ) - { #endif + { setROPumpTargetFlowRateLPM( getTargetFillFlowRateLPM(), TARGET_RO_PRESSURE_PSI ); handleDialysateMixing( getMeasuredROFlowRateLPM() * ML_PER_LITER ); @@ -646,17 +645,16 @@ fillStatus.isThisFirstFill = FALSE; result = DG_FILL_MODE_STATE_PRODUCE_DIALYSATE; -#ifndef DISABLE_ACID_CONDUCTIVITY_TEST } +#ifndef DISABLE_ACID_CONDUCTIVITY_TEST else { // Acid conductivity did not pass, fault SET_ALARM_WITH_1_F32_DATA( ALARM_ID_ACID_CONDUCTIVITY_OUT_OF_RANGE, averageAcidConductivity ); } +#endif } -#endif - return result; }