Index: firmware/App/Modes/ModeFill.c =================================================================== diff -u -r1aeab08c1baf6445514b81fe51fc60a3e536e782 -r3b5080777e3dac1b5802ee18227f66a0b49deefa --- firmware/App/Modes/ModeFill.c (.../ModeFill.c) (revision 1aeab08c1baf6445514b81fe51fc60a3e536e782) +++ firmware/App/Modes/ModeFill.c (.../ModeFill.c) (revision 3b5080777e3dac1b5802ee18227f66a0b49deefa) @@ -323,8 +323,8 @@ { DG_FILL_MODE_STATE_T result = DG_FILL_MODE_STATE_BICARB_PUMP_CHECK; DG_BICARB_CONCENTRATES_RECORD_T bicarb = getBicarbConcentrateCalRecord(); - F32 const measuredROFlowRate_mL_min = getMeasuredROFlowRate() * ML_PER_LITER; - F32 bicarbPumpFlowRate_mL_min = measuredROFlowRate_mL_min * bicarb.bicarbConcentrate[ CAL_DATA_BICARB_CONCENTRATE_1 ].bicarbConcMixRatio + + F32 measuredROFlowRate_mL_min = getMeasuredROFlowRate() * ML_PER_LITER; + F32 bicarbPumpFlowRate_mL_min = measuredROFlowRate_mL_min * bicarb.bicarbConcentrate[ CAL_DATA_BICARB_CONCENTRATE_1 ].bicarbConcMixRatio + CONCENTRATE_PUMP_PRIME_EXTRA_SPEED_ML_MIN; #ifndef DISABLE_DIALYSATE_CHECK F32 const bicarbConductivity = getConductivityValue( CONDUCTIVITYSENSORS_CD2_SENSOR ); @@ -471,11 +471,11 @@ *************************************************************************/ static DG_FILL_MODE_STATE_T handleDeliverDialysateState( void ) { - F32 const measuredROFlowRate_mL_min = getMeasuredROFlowRate() * ML_PER_LITER; - F32 const acidConductivity = getConductivityValue( CONDUCTIVITYSENSORS_CD1_SENSOR ); - F32 const dialysateConductivity = getConductivityValue( CONDUCTIVITYSENSORS_CD2_SENSOR ); - BOOL const isAcidConductivityOutOfRange = ( acidConductivity <= MIN_ACID_CONCENTRATE_CONDUCTIVITY ) || ( acidConductivity >= MAX_ACID_CONCENTRATE_CONDUCTIVITY ); - BOOL const isDialysateConductivityOutOfRange = ( dialysateConductivity <= MIN_DIALYSATE_CONDUCTIVITY ) || ( dialysateConductivity >= MAX_DIALYSATE_CONDUCTIVITY ); + F32 measuredROFlowRate_mL_min = getMeasuredROFlowRate() * ML_PER_LITER; + F32 acidConductivity = getConductivityValue( CONDUCTIVITYSENSORS_CD1_SENSOR ); + F32 dialysateConductivity = getConductivityValue( CONDUCTIVITYSENSORS_CD2_SENSOR ); + BOOL isAcidConductivityOutOfRange = ( acidConductivity <= MIN_ACID_CONCENTRATE_CONDUCTIVITY ) || ( acidConductivity >= MAX_ACID_CONCENTRATE_CONDUCTIVITY ); + BOOL isDialysateConductivityOutOfRange = ( dialysateConductivity <= MIN_DIALYSATE_CONDUCTIVITY ) || ( dialysateConductivity >= MAX_DIALYSATE_CONDUCTIVITY ); DG_FILL_MODE_STATE_T result = DG_FILL_MODE_STATE_DELIVER_DIALYSATE; DG_RESERVOIR_ID_T inactiveReservoir = getInactiveReservoir();