Index: firmware/App/Modes/ModeFill.c =================================================================== diff -u -ra3e3feea8cfb6a1e0137bf09031f0484a3c73332 -rb1b5e8659de8f6109c4d77cda98308d1fd9ef165 --- firmware/App/Modes/ModeFill.c (.../ModeFill.c) (revision a3e3feea8cfb6a1e0137bf09031f0484a3c73332) +++ firmware/App/Modes/ModeFill.c (.../ModeFill.c) (revision b1b5e8659de8f6109c4d77cda98308d1fd9ef165) @@ -209,6 +209,7 @@ if ( TRUE == isPersistentAlarmTriggered( ALARM_ID_DIALYSATE_CONDUCTIVITY_OUT_OF_RANGE, !isDialysateConductivityGood ) ) { +#ifndef DISABLE_DIALYSATE_CHECK if ( FALSE == isAcidConductivityInRange() ) { SET_ALARM_WITH_1_F32_DATA( ALARM_ID_ACID_CONDUCTIVITY_OUT_OF_RANGE, getConductivityValue( CONDUCTIVITYSENSORS_CD1_SENSOR ) ); @@ -217,6 +218,7 @@ { SET_ALARM_WITH_1_F32_DATA( ALARM_ID_DIALYSATE_CONDUCTIVITY_OUT_OF_RANGE, getConductivityValue( CONDUCTIVITYSENSORS_CD2_SENSOR ) ); } +#endif } if ( TRUE == isPersistentAlarmConditionCleared( ALARM_ID_DIALYSATE_CONDUCTIVITY_OUT_OF_RANGE, !isDialysateConductivityGood ) ) @@ -271,7 +273,9 @@ handleDialysateMixing(); +#ifndef DISABLE_DIALYSATE_CHECK if ( TRUE == isBicarbConductivityInRange() ) +#endif { concentrateTestStartTime = getMSTimerCount(); requestConcentratePumpsOn( CONCENTRATEPUMPS_CP1_ACID ); @@ -302,7 +306,9 @@ handleDialysateMixing(); +#ifndef DISABLE_DIALYSATE_CHECK if ( TRUE == isAcidConductivityInRange() ) +#endif { requestConcentratePumpsOn( CONCENTRATEPUMPS_CP1_ACID ); requestConcentratePumpsOn( CONCENTRATEPUMPS_CP2_BICARB );