Index: firmware/App/Modes/ModeFill.c =================================================================== diff -u -rc7ea3969643ead419cbfcd34c3cb203c45bcad71 -re58e3385eaf5e920335f963937b2e9f654a32cfe --- firmware/App/Modes/ModeFill.c (.../ModeFill.c) (revision c7ea3969643ead419cbfcd34c3cb203c45bcad71) +++ firmware/App/Modes/ModeFill.c (.../ModeFill.c) (revision e58e3385eaf5e920335f963937b2e9f654a32cfe) @@ -667,7 +667,7 @@ } #endif - if ( ( TRUE == isConductivityInRange ) || ( TRUE == getTestConfigStatus( TEST_CONFIG_ENABLE_MIXING_WITH_WATER ) ) ) + if ( ( TRUE == isConductivityInRange ) || ( TRUE == getTestConfigStatus( TEST_CONFIG_MIX_WITH_WATER ) ) ) { // Initialization requestConcentratePumpOff( CONCENTRATEPUMPS_CP2_BICARB, NO_PARK_CONC_PUMPS ); @@ -739,7 +739,7 @@ #endif { if ( ( TRUE == isValueWithinPercentRange( averageAcidConductivity, acidCondUSPerCM, FIVE_PERCENT_FACTOR ) ) || - ( TRUE == getTestConfigStatus( TEST_CONFIG_ENABLE_MIXING_WITH_WATER ) ) ) + ( TRUE == getTestConfigStatus( TEST_CONFIG_MIX_WITH_WATER ) ) ) { hasAcidTestPassed = TRUE; } @@ -748,7 +748,7 @@ SET_ALARM_WITH_1_F32_DATA( ALARM_ID_DG_ACID_CONDUCTIVITY_OUT_OF_RANGE, averageAcidConductivity ) } - if ( ( pctDiffInConductivity < FIVE_PERCENT_FACTOR ) || ( TRUE == getTestConfigStatus( TEST_CONFIG_ENABLE_MIXING_WITH_WATER ) ) ) + if ( ( pctDiffInConductivity < FIVE_PERCENT_FACTOR ) || ( TRUE == getTestConfigStatus( TEST_CONFIG_MIX_WITH_WATER ) ) ) { hasCD1CD2TestPassed = TRUE; } @@ -934,7 +934,7 @@ F32 avgCPo = sumFillCPoConductivity / (F32)fillCPoConductivitySampleCnt; // sample count incremented above w/o condition so no need for divide by zero checks F32 avgRR = sumFillRejRatio / (F32)fillCPoConductivitySampleCnt; - if ( ( avgCPo > MAX_CPO_CONDUCTIVITY_ALLOW ) && ( getTestConfigStatus( TEST_CONFIG_ENABLE_MIXING_WITH_WATER ) != FALSE ) ) + if ( ( avgCPo > MAX_CPO_CONDUCTIVITY_ALLOW ) && ( getTestConfigStatus( TEST_CONFIG_MIX_WITH_WATER ) != FALSE ) ) { // Fault alarm per PRS 483 SET_ALARM_WITH_2_F32_DATA( ALARM_ID_DG_OUTLET_PRIMARY_CONDUCTIVITY_OUT_OF_RANGE, avgCPo, MAX_CPO_CONDUCTIVITY_ALLOW ); @@ -965,7 +965,7 @@ { // SRSDG 400 if ( ( FALSE == isValueWithinPercentRange( avgBicarbConductivity, bicarbNormalConductivity, FIVE_PERCENT_FACTOR ) ) && - ( getTestConfigStatus( TEST_CONFIG_ENABLE_MIXING_WITH_WATER ) != TRUE ) ) + ( getTestConfigStatus( TEST_CONFIG_MIX_WITH_WATER ) != TRUE ) ) { setBadAvgConductivityDetectedFlag( TRUE ); // signal idle bad avg conductivity detected setThisFisrtFillFlag( TRUE ); @@ -975,7 +975,7 @@ } if ( ( FALSE == isValueWithinPercentRange( avgAcidConductivity, acidNormalConductivity, FIVE_PERCENT_FACTOR ) ) && - ( getTestConfigStatus( TEST_CONFIG_ENABLE_MIXING_WITH_WATER ) != TRUE ) ) + ( getTestConfigStatus( TEST_CONFIG_MIX_WITH_WATER ) != TRUE ) ) { setBadAvgConductivityDetectedFlag( TRUE ); // signal idle bad avg conductivity detected setThisFisrtFillFlag( TRUE );