Index: firmware/App/Modes/ModeFill.c =================================================================== diff -u -re58e3385eaf5e920335f963937b2e9f654a32cfe -rc940dea020a10f2bab23ae92aa6932c375c26dd4 --- firmware/App/Modes/ModeFill.c (.../ModeFill.c) (revision e58e3385eaf5e920335f963937b2e9f654a32cfe) +++ firmware/App/Modes/ModeFill.c (.../ModeFill.c) (revision c940dea020a10f2bab23ae92aa6932c375c26dd4) @@ -8,7 +8,7 @@ * @file ModeFill.c * * @author (last) Dara Navaei -* @date (last) 26-Apr-2023 +* @date (last) 11-May-2023 * * @author (original) Leonardo Baloa * @date (original) 19-Nov-2019 @@ -609,11 +609,14 @@ bicarbConductivitySampleCount = 0; // Set pumps flow rate to prepare for bicarb conductivity testing - setROPumpTargetFlowRateLPM( RO_PUMP_400_ML_PER_MIN / MILLILITERS_PER_LITER, TARGET_RO_PRESSURE_PSI ); + setROPumpTargetFlowRateLPM( RO_PUMP_800_ML_PER_MIN / MILLILITERS_PER_LITER, TARGET_RO_PRESSURE_PSI ); setConcentratePumpTargetSpeed( CONCENTRATEPUMPS_CP2_BICARB, BICARB_PUMP_40_ML_PER_MIN ); requestConcentratePumpOn( CONCENTRATEPUMPS_CP2_BICARB ); + // Prior to collecting data for bicarb test, change the CD2 calibration table + setCondcutivitySensorCalTable( CONDUCTIVITYSENSORS_CD2_SENSOR, CAL_DATA_CD2_COND_SENSOR_FILL_BICARB_TEST ); + // State transition concentrateTestStartTime = getMSTimerCount(); result = DG_FILL_MODE_STATE_TEST_BICARB_CONDUCTIVITY; @@ -680,6 +683,10 @@ setROPumpTargetFlowRateLPM( RO_PUMP_800_ML_PER_MIN / MILLILITERS_PER_LITER, TARGET_RO_PRESSURE_PSI ); setConcentratePumpTargetSpeed( CONCENTRATEPUMPS_CP1_ACID, ACID_PUMP_20_ML_PER_MIN ); requestConcentratePumpOn( CONCENTRATEPUMPS_CP1_ACID ); + + // Set back the conductivity of CD2 calibration table to the normal calibration table + setCondcutivitySensorCalTable( CONDUCTIVITYSENSORS_CD2_SENSOR, CAL_DATA_CD2_COND_SENSOR ); + // State transition concentrateTestStartTime = getMSTimerCount(); result = DG_FILL_MODE_STATE_TEST_ACID_CONDUCTIVITY;