Index: firmware/App/Modes/ModeHeatDisinfect.c =================================================================== diff -u -r06ccd13d7e04a57dfb750a5d14ae4fea1525d4cc -r75f5cf4a8484f0c08b90edfb4b49a6268e51de65 --- firmware/App/Modes/ModeHeatDisinfect.c (.../ModeHeatDisinfect.c) (revision 06ccd13d7e04a57dfb750a5d14ae4fea1525d4cc) +++ firmware/App/Modes/ModeHeatDisinfect.c (.../ModeHeatDisinfect.c) (revision 75f5cf4a8484f0c08b90edfb4b49a6268e51de65) @@ -91,7 +91,7 @@ // R1 to R2 & R2 to R1 heat disinfect circulation #define HEAT_DISINFECT_TARGET_RO_FLOW_LPM 1.3F ///< Heat disinfect target RO flow rate in L/min. -#define HEAT_DISINFECT_TARGET_RO_FLOW_TRANSFER_LPM 0.8F ///< Heat disinfect target RO flow rate in L/min when transferring between reservoirs. +#define HEAT_DISINFECT_TARGET_RO_FLOW_TRANSFER_LPM 0.5F ///< Heat disinfect target RO flow rate in L/min when transferring between reservoirs. #define HEAT_DISINFECT_MAX_RO_PRESSURE_PSI 30 ///< Heat disinfect maximum RO pressure in psi. #define HEAT_DISINFECT_START_TEMP_TIMEOUT_MS ( 4 * MIN_PER_HOUR * SEC_PER_MIN * MS_PER_SECOND ) ///< Heat disinfect reaching to minimum temperature timeout in milliseconds. #define RSRVRS_TARGET_VOL_OUT_TIMEOUT_MS ( 10 * MS_PER_SECOND ) ///< Reservoirs 1 & 2 maximum volume out of range timeout during heat disinfect. @@ -636,7 +636,6 @@ stateTimer = getMSTimerCount(); stateTrialCounter = 0; state = DG_HEAT_DISINFECT_STATE_FLUSH_CIRCULATION; - } return state;