Index: firmware/App/Modes/ModeHeatDisinfect.c =================================================================== diff -u -r353d01d31bc69d4d1901ff09097ee610bb1c9dbc -rf5dc98b3d5fcd39e3ab502fbe81110e188ce0ec5 --- firmware/App/Modes/ModeHeatDisinfect.c (.../ModeHeatDisinfect.c) (revision 353d01d31bc69d4d1901ff09097ee610bb1c9dbc) +++ firmware/App/Modes/ModeHeatDisinfect.c (.../ModeHeatDisinfect.c) (revision f5dc98b3d5fcd39e3ab502fbe81110e188ce0ec5) @@ -607,9 +607,8 @@ if ( TRUE == didTimeout( stateTimer, FLUSH_DRAIN_WAIT_TIME_MS ) ) { // If the inlet temperature and conductivity are in range, move onto the next state - //if ( ( getTemperatureValue( TEMPSENSORS_INLET_PRIMARY_HEATER ) > MIN_INLET_TEMPERATURE_C ) && - // ( getConductivityValue( CONDUCTIVITYSENSORS_CPI_SENSOR ) <= MAX_INLET_CONDUCTIVITY_US_PER_CM ) ) - if ( TRUE ) // DEBUG_DENALI enable the conductivity check + if ( ( getTemperatureValue( TEMPSENSORS_INLET_PRIMARY_HEATER ) > MIN_INLET_TEMPERATURE_C ) && + ( getConductivityValue( CONDUCTIVITYSENSORS_CPI_SENSOR ) <= MAX_INLET_CONDUCTIVITY_US_PER_CM ) ) { setValveState( VPD, VALVE_STATE_OPEN_C_TO_NC ); setROPumpTargetFlowRateLPM( RO_PUMP_TARGET_FLUSH_FILL_FLOW_RATE_LPM, MAX_RO_PUMP_FLUSH_FILL_PRESSURE_PSI );