Index: firmware/App/Modes/ModeHeatDisinfect.c =================================================================== diff -u -r3e5c16b70fb34da7d58f5cf6efff896fdb3d00c3 -ra33ab8f72a09be4650026964dd2bd35ff95c8431 --- firmware/App/Modes/ModeHeatDisinfect.c (.../ModeHeatDisinfect.c) (revision 3e5c16b70fb34da7d58f5cf6efff896fdb3d00c3) +++ firmware/App/Modes/ModeHeatDisinfect.c (.../ModeHeatDisinfect.c) (revision a33ab8f72a09be4650026964dd2bd35ff95c8431) @@ -79,6 +79,7 @@ #define RSRVRS_DRAIN_TIMEOUT_MS ( 2 * SEC_PER_MIN * MS_PER_SECOND ) ///< Reservoirs 1 & 2 drain timeout in ms. // Target temperature +#define HEAT_DISINFECT_PRIM_HEATER_FLUSH_TARGET_TEMP_C 60.0F ///< Heat disinfect primary heater flush target temperature C. #define HEAT_DISINFECT_PRIM_HEATER_TARGET_TEMP_C 90.0F ///< Heat disinfect primary heater target temperature in C. #define HEAT_DISINFECT_TRIM_HEATER_TARGET_TEMP_C 90.0F ///< Heat disinfect trimmer heater target temperature in C. #define HEAT_DISINFECT_CONC_PUMPS_START_TEMP_C 75.0F ///< Heat disinfect start concentrate pumps target temperature in C. @@ -724,7 +725,7 @@ if ( nelsonSupport != NELSON_INOCULATE ) #endif { - setHeaterTargetTemperature( DG_PRIMARY_HEATER, HEAT_DISINFECT_PRIM_HEATER_TARGET_TEMP_C ); + setHeaterTargetTemperature( DG_PRIMARY_HEATER, HEAT_DISINFECT_PRIM_HEATER_FLUSH_TARGET_TEMP_C ); startHeater( DG_PRIMARY_HEATER ); }