Index: firmware/App/Modes/ModeHeatDisinfect.c =================================================================== diff -u -r922516483829939a2a387d4a2fddeccdb4c454d4 -r54e58f64179ea382d2e2c403c8c3b9a15a612636 --- firmware/App/Modes/ModeHeatDisinfect.c (.../ModeHeatDisinfect.c) (revision 922516483829939a2a387d4a2fddeccdb4c454d4) +++ firmware/App/Modes/ModeHeatDisinfect.c (.../ModeHeatDisinfect.c) (revision 54e58f64179ea382d2e2c403c8c3b9a15a612636) @@ -447,13 +447,8 @@ setValveState( VPI, VALVE_STATE_CLOSED ); // Request a tare for reservoir 1 tareReservoir(); -#ifndef V_2_SYSTEM // Set the actuators to drain R1 setValveState( VRD1, VALVE_STATE_OPEN ); -#else - // Set the actuators to drain R1 - setValveState( VRD, VALVE_STATE_R1_C_TO_NC ); -#endif setDrainPumpTargetRPM( DRAIN_PUMP_TARGET_RPM ); rsrvrFillStableTimeCounter = 0; @@ -492,12 +487,8 @@ // Set the valves to flush the recirculation line setValveState( VPI, VALVE_STATE_OPEN ); -#ifndef V_2_SYSTEM setValveState( VPD, VALVE_STATE_OPEN_C_TO_NC ); setValveState( VRD1, VALVE_STATE_CLOSED ); -#else - setValveState( VPD, VALVE_STATE_OPEN_C_TO_NO ); -#endif setValveState( VPO, VALVE_STATE_NOFILL_C_TO_NO ); setValveState( VDR, VALVE_STATE_DRAIN_C_TO_NO ); setValveState( VRC, VALVE_STATE_DRAIN_C_TO_NO ); @@ -518,17 +509,11 @@ // Request a tare for reservoir 2 tareReservoir(); -#ifndef V_2_SYSTEM // Done with draining R1, close it setValveState( VRD1, VALVE_STATE_CLOSED ); // Set the actuators to drain R2. // NOTE: Drain pump is already on and VDr is already on drain state setValveState( VRD2, VALVE_STATE_OPEN ); -#else - // Set the actuators to drain R2. - // NOTE: Drain pump is already on and VDr is already on drain state - setValveState( VRD, VALVE_STATE_R2_C_TO_NO ); -#endif state = DG_HEAT_DISINFECT_STATE_DRAIN_R2; } @@ -567,23 +552,17 @@ { if ( TRUE == isThisLastDrain ) { -#ifndef V_2_SYSTEM setValveState( VRD1, VALVE_STATE_OPEN ); setValveState( VRD2, VALVE_STATE_CLOSED ); -#else - setValveState( VRD, VALVE_STATE_R1_C_TO_NC ); -#endif rsrvr1Status = DG_RESERVOIR_ABOVE_TARGET; state = DG_HEAT_DISINFECT_STATE_DRAIN_R1; } else { signalDrainPumpHardStop(); -#ifndef V_2_SYSTEM // Done with draining R2, close it setValveState( VRD2, VALVE_STATE_CLOSED ); -#endif setValveState( VPI, VALVE_STATE_OPEN ); stateTrialCounter = 0; stateTimer = getMSTimerCount(); @@ -623,11 +602,7 @@ if ( ( getTemperatureValue( TEMPSENSORS_INLET_PRIMARY_HEATER ) > MIN_INLET_TEMPERATURE_C ) && ( getConductivityValue( CONDUCTIVITYSENSORS_CPI_SENSOR ) <= MAX_INLET_CONDUCTIVITY_US_PER_CM ) ) { -#ifndef V_2_SYSTEM setValveState( VPD, VALVE_STATE_OPEN_C_TO_NC ); -#else - setValveState( VPD, VALVE_STATE_OPEN_C_TO_NO ); -#endif setROPumpTargetFlowRate( RO_PUMP_TARGET_FLUSH_FILL_FLOW_RATE_LPM, MAX_RO_PUMP_FLUSH_FILL_PRESSURE_PSI ); stateTimer = getMSTimerCount(); stateTrialCounter = 0; @@ -725,8 +700,8 @@ rsrvr2Status = DG_RESERVOIR_BELOW_TARGET; // Done with flushing the concentrate pumps line - requestConcentratePumpsOff( CONCENTRATEPUMPS_CP1_ACID ); - requestConcentratePumpsOff( CONCENTRATEPUMPS_CP2_BICARB ); + requestConcentratePumpOff( CONCENTRATEPUMPS_CP1_ACID ); + requestConcentratePumpOff( CONCENTRATEPUMPS_CP2_BICARB ); setValveState( VPO, VALVE_STATE_FILL_C_TO_NC ); setValveState( VRF, VALVE_STATE_R1_C_TO_NC ); @@ -784,11 +759,7 @@ setValveState( VRF, VALVE_STATE_R2_C_TO_NO ); setValveState( VRI, VALVE_STATE_R1_C_TO_NO ); setValveState( VRO, VALVE_STATE_R2_C_TO_NC ); -#ifndef V_2_SYSTEM setValveState( VRD1, VALVE_STATE_OPEN ); -#else - setValveState( VRD, VALVE_STATE_R1_C_TO_NC ); -#endif setDrainPumpTargetRPM( DRAIN_PUMP_TARGET_RPM ); stateTimer = getMSTimerCount(); @@ -838,10 +809,8 @@ { // Done with draining R1 signalDrainPumpHardStop(); -#ifndef V_2_SYSTEM // Close VRD1 setValveState( VRD1, VALVE_STATE_CLOSED ); -#endif } else if ( DG_RESERVOIR_NOT_REACHED_TARGET == rsrvr1Status ) { @@ -882,18 +851,10 @@ // Set the valves to drain R2 and no fill setValveState( VPI, VALVE_STATE_CLOSED ); -#ifndef V_2_SYSTEM setValveState( VPD, VALVE_STATE_OPEN_C_TO_NC ); -#else - setValveState( VPD, VALVE_STATE_OPEN_C_TO_NO ); -#endif setValveState( VPO, VALVE_STATE_NOFILL_C_TO_NO ); setValveState( VRO, VALVE_STATE_R1_C_TO_NO ); -#ifndef V_2_SYSTEM setValveState( VRD2, VALVE_STATE_OPEN ); -#else - setValveState( VRD, VALVE_STATE_R2_C_TO_NO ); -#endif setDrainPumpTargetRPM( DRAIN_PUMP_TARGET_RPM ); // Start the timer for drain timeout stateTimer = getMSTimerCount(); @@ -937,12 +898,8 @@ } else if ( DG_RESERVOIR_REACHED_TARGET == rsrvr2Status ) { -#ifndef V_2_SYSTEM setValveState( VRD1, VALVE_STATE_OPEN ); setValveState( VRD2, VALVE_STATE_CLOSED ); -#else - setValveState( VRD, VALVE_STATE_R1_C_TO_NC ); -#endif // Start the timer for drain timeout stateTimer = getMSTimerCount(); rsrvr1Status = DG_RESERVOIR_ABOVE_TARGET; @@ -984,28 +941,20 @@ // Prepare for filling the reservoirs and heating the water setValveState( VPI, VALVE_STATE_OPEN ); -#ifndef V_2_SYSTEM setValveState( VPD, VALVE_STATE_OPEN_C_TO_NC ); -#else - setValveState( VPD, VALVE_STATE_OPEN_C_TO_NO ); -#endif setValveState( VPO, VALVE_STATE_FILL_C_TO_NC ); setValveState( VRO, VALVE_STATE_R1_C_TO_NO ); setValveState( VRI, VALVE_STATE_R2_C_TO_NC ); setValveState( VRF, VALVE_STATE_R1_C_TO_NC ); -#ifndef V_2_SYSTEM setValveState( VRD1, VALVE_STATE_CLOSED ); setValveState( VRD2, VALVE_STATE_OPEN ); -#else - setValveState( VRD, VALVE_STATE_R2_C_TO_NO ); -#endif // Turn on the RO pump setROPumpTargetFlowRate( RO_PUMP_TARGET_FLUSH_FILL_FLOW_RATE_LPM, MAX_RO_PUMP_FLUSH_FILL_PRESSURE_PSI ); // Start heating the water while we are filling up the reservoirs - setPrimaryHeaterTargetTemperature( HEAT_DISINFECT_TARGET_TEMPERATURE_C ); - startPrimaryHeater(); + setHeaterTargetTemperature( DG_PRIMARY_HEATER, HEAT_DISINFECT_TARGET_TEMPERATURE_C ); + startHeater( DG_PRIMARY_HEATER ); rsrvr1Status = DG_RESERVOIR_BELOW_TARGET; rsrvr2Status = DG_RESERVOIR_BELOW_TARGET; @@ -1060,11 +1009,7 @@ // Set the valves to drain R2 and no fill setValveState( VPI, VALVE_STATE_CLOSED ); setValveState( VBF, VALVE_STATE_OPEN ); -#ifndef V_2_SYSTEM setValveState( VPD, VALVE_STATE_DRAIN_C_TO_NO ); -#else - setValveState( VPD, VALVE_STATE_DRAIN_C_TO_NC ); -#endif setValveState( VDR, VALVE_STATE_RECIRC_C_TO_NC ); setValveState( VRC, VALVE_STATE_RECIRC_C_TO_NC ); @@ -1076,8 +1021,8 @@ setROPumpTargetFlowRate( HEAT_DISINFECT_TARGET_RO_FLOW_LPM, HEAT_DISINFECT_MAX_RO_PRESSURE_PSI ); // Start the trimmer heater since we are recirculating water and there is flow in the shunt line - setTrimmerHeaterTargetTemperature( HEAT_DISINFECT_TARGET_TEMPERATURE_C ); - startTrimmerHeater(); + setHeaterTargetTemperature( DG_TRIMMER_HEATER, HEAT_DISINFECT_TARGET_TEMPERATURE_C ); + startHeater( DG_TRIMMER_HEATER ); // Get the current volumes of R1 & R2. These values will be used to make sure the reservoirs' // volume does not change more than a certain amount during the actual heat disinfect cycle @@ -1133,16 +1078,12 @@ case HEAT_DISINFECT_COMPLETE: - requestConcentratePumpsOff( CONCENTRATEPUMPS_CP1_ACID ); - requestConcentratePumpsOff( CONCENTRATEPUMPS_CP2_BICARB ); + requestConcentratePumpOff( CONCENTRATEPUMPS_CP1_ACID ); + requestConcentratePumpOff( CONCENTRATEPUMPS_CP2_BICARB ); // Set the valves to transfer hot water from R1 to R2 and fill up R2. setValveState( VRO, VALVE_STATE_R2_C_TO_NC ); -#ifndef V_2_SYSTEM setValveState( VRD1, VALVE_STATE_OPEN ); setValveState( VRD2, VALVE_STATE_CLOSED ); -#else - setValveState( VRD, VALVE_STATE_R1_C_TO_NC ); -#endif setValveState( VRI, VALVE_STATE_R1_C_TO_NO ); setValveState( VRF, VALVE_STATE_R2_C_TO_NO ); // Although there is fluid in both reservoirs, but they are set to empty @@ -1241,8 +1182,8 @@ case HEAT_DISINFECT_COMPLETE: // Turn off the heaters - stopPrimaryHeater(); - stopTrimmerHeater(); + stopHeater( DG_PRIMARY_HEATER ); + stopHeater( DG_TRIMMER_HEATER ); stateTimer = getMSTimerCount(); state = DG_HEAT_DISINFECT_STATE_COOL_DOWN_HEATERS; break; @@ -1285,12 +1226,8 @@ setValveState( VPI, VALVE_STATE_CLOSED ); setValveState( VBF, VALVE_STATE_CLOSED ); setValveState( VRO, VALVE_STATE_R1_C_TO_NO ); -#ifndef V_2_SYSTEM setValveState( VPD, VALVE_STATE_DRAIN_C_TO_NO ); setValveState( VRD1, VALVE_STATE_CLOSED ); -#else - setValveState( VPD, VALVE_STATE_DRAIN_C_TO_NC ); -#endif ROFCoolingTimer = getMSTimerCount(); stateTimer = getMSTimerCount(); @@ -1392,9 +1329,7 @@ { isDrainPumpInMixDrainOn = TRUE; -#ifndef V_2_SYSTEM setValveState( VRD1, VALVE_STATE_OPEN ); -#endif // Turn on the drain pump to drain the reservoirs in open loop mode setDrainPumpTargetRPM( DRAIN_PUMP_RPM_IN_MIX_DRAIN ); } @@ -1406,15 +1341,10 @@ } else if ( DG_RESERVOIR_REACHED_TARGET == rsrvr1Status ) { -#ifndef V_2_SYSTEM // Done with draining reservoir 1 setValveState( VRD1, VALVE_STATE_CLOSED ); // Set the drain valve to reservoir 2 setValveState( VRD2, VALVE_STATE_OPEN ); -#else - // Set the drain valve to reservoir 2 - setValveState( VRD, VALVE_STATE_R2_C_TO_NO ); -#endif rsrvr2Status = DG_RESERVOIR_ABOVE_TARGET; stateTimer = getMSTimerCount(); @@ -1455,13 +1385,9 @@ // Set the valves to fill up R1 and overflow to R2 setValveState( VPI, VALVE_STATE_OPEN ); -#ifndef V_2_SYSTEM setValveState( VPD, VALVE_STATE_OPEN_C_TO_NC ); // Done with draining reservoir 2 setValveState( VRD2, VALVE_STATE_CLOSED ); -#else - setValveState( VPD, VALVE_STATE_OPEN_C_TO_NO ); -#endif setValveState( VPO, VALVE_STATE_FILL_C_TO_NC ); setValveState( VDR, VALVE_STATE_DRAIN_C_TO_NO ); setValveState( VRC, VALVE_STATE_DRAIN_C_TO_NO ); @@ -1524,11 +1450,7 @@ // Set the valves to rinse R2 to R1 and drain R1 setValveState( VRF, VALVE_STATE_R2_C_TO_NO ); setValveState( VRO, VALVE_STATE_R2_C_TO_NC ); -#ifndef V_2_SYSTEM setValveState( VRD1, VALVE_STATE_OPEN ); -#else - setValveState( VRD, VALVE_STATE_R1_C_TO_NC ); -#endif setValveState( VRI, VALVE_STATE_R1_C_TO_NO ); setDrainPumpTargetRPM( DRAIN_PUMP_TARGET_RPM ); @@ -1577,9 +1499,7 @@ { // Done with draining R1 signalDrainPumpHardStop(); -#ifndef V_2_SYSTEM setValveState( VRD1, VALVE_STATE_CLOSED ); -#endif } else if ( DG_RESERVOIR_NOT_REACHED_TARGET == rsrvr1Status ) { @@ -1619,18 +1539,10 @@ // De-energize all the valves and set the VDr to drain R2 setValveState( VPI, VALVE_STATE_CLOSED ); -#ifndef V_2_SYSTEM setValveState( VPD, VALVE_STATE_OPEN_C_TO_NC ); -#else - setValveState( VPD, VALVE_STATE_OPEN_C_TO_NO ); -#endif setValveState( VPO, VALVE_STATE_NOFILL_C_TO_NO ); setValveState( VRO, VALVE_STATE_R1_C_TO_NO ); -#ifndef V_2_SYSTEM setValveState( VRD2, VALVE_STATE_OPEN ); -#else - setValveState( VRD, VALVE_STATE_R2_C_TO_NO ); -#endif setValveState( VRI, VALVE_STATE_R1_C_TO_NO ); setValveState( VRF, VALVE_STATE_R2_C_TO_NO ); @@ -1756,11 +1668,7 @@ { // The fluid is hot so this is a mix drain. Set the VPd to direct the cold inlet fluid to drain setValveState( VPI, VALVE_STATE_OPEN ); -#ifndef V_2_SYSTEM setValveState( VPD, VALVE_STATE_DRAIN_C_TO_NO ); -#else - setValveState( VPD, VALVE_STATE_DRAIN_C_TO_NC ); -#endif targetRPM = DRAIN_PUMP_RPM_IN_MIX_DRAIN; cancellationMode = CANCELLATION_MODE_HOT; @@ -1787,14 +1695,9 @@ if ( DG_RESERVOIR_REACHED_TARGET == rsrvr2Status ) { -#ifndef V_2_SYSTEM // Set the drain valve to reservoir 1 and close reservoir 2 setValveState( VRD1, VALVE_STATE_OPEN ); setValveState( VRD2, VALVE_STATE_CLOSED ); -#else - // Set the drain valve to reservoir 1 - setValveState( VRD, VALVE_STATE_R1_C_TO_NC ); -#endif } } else if ( DG_RESERVOIR_NOT_REACHED_TARGET == rsrvr2Status )