Index: firmware/App/Modes/ModeHeatDisinfect.c =================================================================== diff -u -rc3dad292358c336c5b37716292f84d79b9881e2f -rc9a2e913b21980bb44a5418f6acf01a7fa77a9c9 --- firmware/App/Modes/ModeHeatDisinfect.c (.../ModeHeatDisinfect.c) (revision c3dad292358c336c5b37716292f84d79b9881e2f) +++ firmware/App/Modes/ModeHeatDisinfect.c (.../ModeHeatDisinfect.c) (revision c9a2e913b21980bb44a5418f6acf01a7fa77a9c9) @@ -622,7 +622,7 @@ // Couldn't get a good water sample after a couple of trials and the disinfect cycle failed else { - alarmDetectedPendingTrigger = ALARM_ID_DG_HEAT_DISINFECT_INLET_COND_OR_TEMP_OUT; + alarmDetectedPendingTrigger = ALARM_ID_DG_HEAT_DISINFECT_INLET_COND_AND_TEMP_OUT; prevHeatDisinfectState = state; state = DG_HEAT_DISINFECT_STATE_CANCEL_BASIC_PATH; }