Index: firmware/App/Modes/ModeHeatDisinfectActiveCool.c =================================================================== diff -u -rc230be1bd4296324bf5dfc288c212eb7c2ce5d2d -rff676a9640e5dfe162de78d2b24f356f63bad5bb --- firmware/App/Modes/ModeHeatDisinfectActiveCool.c (.../ModeHeatDisinfectActiveCool.c) (revision c230be1bd4296324bf5dfc288c212eb7c2ce5d2d) +++ firmware/App/Modes/ModeHeatDisinfectActiveCool.c (.../ModeHeatDisinfectActiveCool.c) (revision ff676a9640e5dfe162de78d2b24f356f63bad5bb) @@ -723,7 +723,7 @@ { SET_ALARM_WITH_1_U32_DATA( alarmDetectedPendingTrigger, heatDisinfectActiceCoolPrevState ) } - requestNewOperationMode( DG_MODE_FAUL ); + requestNewOperationMode( DG_MODE_STAN ); } /*********************************************************************//** @@ -771,10 +771,14 @@ { // Set the variables to fail and go to cancel water path. Set the pending alarm to no alarm so the cancel water path // will not be raising the alarm at end of the cancel water path. The recoverable alarm is raised here in this function - U32 cap = (U32)( STATE_OPEN == getSwitchStatus( CONCENTRATE_CAP ) ? CONCENTRATE_CAP : DIALYSATE_CAP ); heatDisinfectActiceCoolPrevState = heatDisinfectActiveCoolState; heatDisinfectActiveCoolState = DG_HEAT_DISINFECT_ACTIVE_COOL_CANCEL_BASIC_PATH_STATE; - alarmDetectedPendingTrigger = ALARM_ID_DG_DIALYSATE_OR_CONC_CAP_NOT_IN_PROPER_POSITION; + alarmDetectedPendingTrigger = ALARM_ID_DG_DIALYSATE_CAP_NOT_IN_PROPER_POSITION; + + if ( STATE_OPEN == getSwitchStatus( CONCENTRATE_CAP ) ) + { + alarmDetectedPendingTrigger = ALARM_ID_DG_CONCENTRATE_CAP_NOT_IN_PROPER_POSITION; + } } }