Index: firmware/App/Controllers/Pressures.c =================================================================== diff -u -r4fbaa823eb08d369ea24d4ce59e77b6f2ad8e97a -r49a77f981e17f8708c8bdbec7dcc2fd25c5510fd --- firmware/App/Controllers/Pressures.c (.../Pressures.c) (revision 4fbaa823eb08d369ea24d4ce59e77b6f2ad8e97a) +++ firmware/App/Controllers/Pressures.c (.../Pressures.c) (revision 49a77f981e17f8708c8bdbec7dcc2fd25c5510fd) @@ -56,13 +56,13 @@ #define MIN_VALID_BARO_PRESSURE_PSIA 10.1F ///< Minimum valid barometric pressure in psia. #define MAX_VALID_BARO_PRESSURE_PSIA 15.4F ///< Maximum valid barometric pressure in psia. -#define MAX_INLET_WATER_PRESSURE_WARNING_LOW_PSIG 14.0F ///< Maximum allowed low pressure value in psig. +#define MAX_INLET_WATER_PRESSURE_WARNING_LOW_PSIG 23.0F ///< Maximum allowed low pressure value in psig. #define MIN_INLET_WATER_PRESSURE_WARNING_LOW_PSIG 30.0F ///< Minimum allowed low pressure value in psig. #define MAX_INLET_WATER_PRESSURE_WARNING_HIGH_PSIG 80.0F ///< Maximum allowed high pressure value in psig. #define MIN_INLET_WATER_PRESSURE_WARNING_HIGH_PSIG 78.0F ///< Minimum allowed high pressure value in psig. -#define INLET_WATER_PRES_OUT_OF_RANGE_TIMEOUT_MS ( 30 * MS_PER_SECOND ) ///< Persistence period for pressure out of range error in milliseconds. +#define INLET_WATER_PRES_OUT_OF_RANGE_TIMEOUT_MS ( 20 * MS_PER_SECOND ) ///< Persistence period for pressure out of range error in milliseconds. #define INLET_WATER_PRES_OUT_OF_RANGE_CLEAR_MS ( 1 * MS_PER_SECOND ) ///< Persistence period for pressure out of range clear in milliseconds. #define PRESSURE_OUT_OF_RANGE_TIMEOUT_MS ( 5 * MS_PER_SECOND ) ///< Pressure out of range persistence period in milliseconds. #define PRESSURES_DATA_PUB_INTERVAL ( MS_PER_SECOND / TASK_PRIORITY_INTERVAL ) ///< Interval (ms/task time) at which the pressures data is published on the CAN bus.