Index: firmware/App/Modes/ModeHeatDisinfect.c =================================================================== diff -u -r44cc25461e4936f7928904640c87a03ef7cf3c16 -ra58aa4a92a57c8b7dbec31a2f834916dc4612e4b --- firmware/App/Modes/ModeHeatDisinfect.c (.../ModeHeatDisinfect.c) (revision 44cc25461e4936f7928904640c87a03ef7cf3c16) +++ firmware/App/Modes/ModeHeatDisinfect.c (.../ModeHeatDisinfect.c) (revision a58aa4a92a57c8b7dbec31a2f834916dc4612e4b) @@ -130,7 +130,7 @@ static BOOL isTemperatureInRange( void ); static void execHeatDisinfectInternalStates( void ); -static void publishHeatDisinfectData ( void ); +static void publishHeatDisinfectData( void ); static U32 getPublishHeatDisinfectDataInterval( void ); static DATA_GET_PROTOTYPE( U32, getPublishHeatDisinfectDataInterval ); static DATA_GET_PROTOTYPE( F32, getRecirculationDuration ); @@ -186,7 +186,7 @@ heatDisinfectState = handleHeatDisinfectStart(); break; - case DG_HEAT_DISINFECT_EVACUATE_DIALYSATE_FILL_WITH_WATER: + case DG_HEAT_DISINFECT_STATE_EVACUATE_DIALYSATE_FILL_WITH_WATER: heatDisinfectState = handleHeatDisinfectEvacuateDialysateFillWithWater(); break; @@ -279,7 +279,7 @@ static DG_HEAT_DISINFECT_STATE_T handleHeatDisinfectStart( void ) { - DG_HEAT_DISINFECT_STATE_T state = DG_HEAT_DISINFECT_EVACUATE_DIALYSATE_FILL_WITH_WATER; + DG_HEAT_DISINFECT_STATE_T state = DG_HEAT_DISINFECT_STATE_EVACUATE_DIALYSATE_FILL_WITH_WATER; stopActuators(); drainPumpTargetRPM = DRAIN_PUMP_EVACUATE_FLUID_TARGET_RPM; @@ -292,7 +292,7 @@ static DG_HEAT_DISINFECT_STATE_T handleHeatDisinfectEvacuateDialysateFillWithWater( void ) { - DG_HEAT_DISINFECT_STATE_T state = DG_HEAT_DISINFECT_EVACUATE_DIALYSATE_FILL_WITH_WATER; + DG_HEAT_DISINFECT_STATE_T state = DG_HEAT_DISINFECT_STATE_EVACUATE_DIALYSATE_FILL_WITH_WATER; execHeatDisinfectInternalStates(); //TODO test to see if it is needed to be in an if statement @@ -547,7 +547,7 @@ INTERNAL_HEAT_DISINFECT_STATE_T state = INTERNAL_HEAT_DISINFECT_STATE_OFF; // If the state is evacuate dialysate, start with evacuating recirculate path - if ( heatDisinfectState == DG_HEAT_DISINFECT_EVACUATE_DIALYSATE_FILL_WITH_WATER ) + if ( heatDisinfectState == DG_HEAT_DISINFECT_STATE_EVACUATE_DIALYSATE_FILL_WITH_WATER ) { // Set the actuators for evacuate recirculation path setValveState ( VPI, VALVE_STATE_OPEN ); @@ -611,7 +611,7 @@ if ( reservoir1Weight >= FULL_RESERVOIRS_WEIGHT_GRAMS && reservoir2Weight >= FULL_RESERVOIRS_WEIGHT_GRAMS ) { - if ( heatDisinfectState == DG_HEAT_DISINFECT_EVACUATE_DIALYSATE_FILL_WITH_WATER ) + if ( heatDisinfectState == DG_HEAT_DISINFECT_STATE_EVACUATE_DIALYSATE_FILL_WITH_WATER ) { stopActuators(); state = INTERNAL_HEAT_DISINFECT_STATE_COMPLETE; @@ -703,7 +703,7 @@ if ( reservoir2Weight <= EMPTY_RESERVOIRS_WEIGHT_GRAMS ) { - if ( heatDisinfectState == DG_HEAT_DISINFECT_EVACUATE_DIALYSATE_FILL_WITH_WATER ) + if ( heatDisinfectState == DG_HEAT_DISINFECT_STATE_EVACUATE_DIALYSATE_FILL_WITH_WATER ) { setActuatorsToFillWater(); state = INTERNAL_HEAT_DISINFECT_STATE_FILL_WITH_WATER;