Index: firmware/App/Controllers/ConcentratePumps.c =================================================================== diff -u -r1cdf34a32afca9bbfd8731d6884f66b330dd97c9 -raf181317c8fc656d5be1d3b9444d81c8c7f154b3 --- firmware/App/Controllers/ConcentratePumps.c (.../ConcentratePumps.c) (revision 1cdf34a32afca9bbfd8731d6884f66b330dd97c9) +++ firmware/App/Controllers/ConcentratePumps.c (.../ConcentratePumps.c) (revision af181317c8fc656d5be1d3b9444d81c8c7f154b3) @@ -175,8 +175,8 @@ stopConcentratePump( pumpId ); } - initPersistentAlarm( ALARM_ID_DG_CP1_SPEED_CONTROL_ERROR, CONCENTRATE_PUMP_SPEED_OUT_OF_RANGE_TIMEOUT_MS, CONCENTRATE_PUMP_SPEED_OUT_OF_RANGE_TIMEOUT_MS ); - initPersistentAlarm( ALARM_ID_DG_CP2_SPEED_CONTROL_ERROR, CONCENTRATE_PUMP_SPEED_OUT_OF_RANGE_TIMEOUT_MS, CONCENTRATE_PUMP_SPEED_OUT_OF_RANGE_TIMEOUT_MS ); + initPersistentAlarm( ALARM_ID_DG_CP1_SPEED_CONTROL_ERROR, 0, CONCENTRATE_PUMP_SPEED_OUT_OF_RANGE_TIMEOUT_MS ); + initPersistentAlarm( ALARM_ID_DG_CP2_SPEED_CONTROL_ERROR, 0, CONCENTRATE_PUMP_SPEED_OUT_OF_RANGE_TIMEOUT_MS ); initPersistentAlarm( ALARM_ID_DG_CONC_PUMP_HALL_SENSOR_OUT_OF_RANGE, CONCENTRATE_PUMP_HALL_SENSORS_OUT_OF_RANGE_TIME_MS, CONCENTRATE_PUMP_HALL_SENSORS_OUT_OF_RANGE_TIME_MS ); initPersistentAlarm( ALARM_ID_DG_ACID_CONCENTRATE_PUMP_PARK_FAULT, CONENTREATE_PUMP_PARK_FAULT_TIMEOUT_MS, CONENTREATE_PUMP_PARK_FAULT_TIMEOUT_MS ); initPersistentAlarm( ALARM_ID_DG_BICARB_CONCENTRATE_PUMP_PARK_FAULT, CONENTREATE_PUMP_PARK_FAULT_TIMEOUT_MS, CONENTREATE_PUMP_PARK_FAULT_TIMEOUT_MS );