Index: firmware/App/Modes/ModeHeatDisinfect.c =================================================================== diff -u -r56ba1b163b0cbf6953638065b2108f745b17ec8f -rfe036771b201ca7320c90ea994ce218dd3196f3c --- firmware/App/Modes/ModeHeatDisinfect.c (.../ModeHeatDisinfect.c) (revision 56ba1b163b0cbf6953638065b2108f745b17ec8f) +++ firmware/App/Modes/ModeHeatDisinfect.c (.../ModeHeatDisinfect.c) (revision fe036771b201ca7320c90ea994ce218dd3196f3c) @@ -113,7 +113,7 @@ #ifndef _RELEASE_ #define NELSON_SUPPORT_TARGET_TEMP_C 6.0F // Nelson support heat disinfect target temperature in C. #define NELSON_SUPPORT_STOP_TEMP_C ( NELSON_SUPPORT_TARGET_TEMP_C - 5.0F ) // Nelson support heat disinfect stop temperature in C. -#define NELSON_SUPPORT_DISINFECT_TIME_MS ( 2 * MIN_PER_HOUR * SEC_PER_MIN * MS_PER_SECOND ) // Nelson support heat disinfect time in milliseconds. +#define NELSON_SUPPORT_DISINFECT_TIME_MS ( 5 * SEC_PER_MIN * MS_PER_SECOND ) //( 2 * MIN_PER_HOUR * SEC_PER_MIN * MS_PER_SECOND ) // Nelson support heat disinfect time in milliseconds. #endif /// Cancellation paths @@ -1049,6 +1049,8 @@ setValveState( VPD, VALVE_STATE_DRAIN_C_TO_NO ); setValveState( VDR, VALVE_STATE_RECIRC_C_TO_NC ); setValveState( VRC, VALVE_STATE_RECIRC_C_TO_NC ); + setValveState( VRD1, VALVE_STATE_CLOSED ); + setValveState( VRD2, VALVE_STATE_OPEN ); signalROPumpHardStop(); // Set the drain pump to control mode setDrainPumpTargetOutletFlowLPM( HEAT_DISINFECT_TARGET_RO_FLOW_LPM );