Index: dialin/hd/treatment.py =================================================================== diff -u -r2e392c92d55178f457a67423ba8c503a86dcf3c8 -r2b80b3c2a26012faf121aeb6be0b3f983061be1c --- dialin/hd/treatment.py (.../treatment.py) (revision 2e392c92d55178f457a67423ba8c503a86dcf3c8) +++ dialin/hd/treatment.py (.../treatment.py) (revision 2b80b3c2a26012faf121aeb6be0b3f983061be1c) @@ -782,7 +782,7 @@ Constructs and sends the set arterial pressure lower alarm limit parameter command. Constraints: Must be logged into HD. - Pressure must be integer and should be between -300 and +200 mmHg + Pressure must be integer and should be between -300 and +0 mmHg @param pres: integer - set arterial pressure lower alarm limit (in mmHg) @return: 1 if successful, zero otherwise @@ -818,7 +818,7 @@ Constructs and sends the set arterial pressure upper alarm limit parameter command. Constraints: Must be logged into HD. - Pressure must be integer and should be between -300 and +200 mmHg + Pressure must be integer and should be between -300 and +0 mmHg @param pres: integer - set arterial pressure upper alarm limit (in mmHg) @return: 1 if successful, zero otherwise @@ -854,7 +854,7 @@ Constructs and sends the set venous pressure lower alarm limit parameter command. Constraints: Must be logged into HD. - Pressure must be integer and should be between -100 and +600 mmHg + Pressure must be integer and should be between +20 and +600 mmHg @param pres: integer - set venous pressure lower alarm limit (in mmHg) @return: 1 if successful, zero otherwise @@ -890,7 +890,7 @@ Constructs and sends the set venous pressure upper alarm limit parameter command. Constraints: Must be logged into HD. - Pressure must be integer and should be between -100 and +600 mmHg + Pressure must be integer and should be between +20 and +600 mmHg @param pres: integer - set venous pressure upper alarm limit (in mmHg) @return: 1 if successful, zero otherwise