Index: dialin/common/hd_defs.py =================================================================== diff -u -rbc39c08a5a630d797c3358f1b58c720e02bbdbf1 -rda7d431511b1fa41d77afa63d538931c13f69775 --- dialin/common/hd_defs.py (.../hd_defs.py) (revision bc39c08a5a630d797c3358f1b58c720e02bbdbf1) +++ dialin/common/hd_defs.py (.../hd_defs.py) (revision da7d431511b1fa41d77afa63d538931c13f69775) @@ -49,32 +49,31 @@ @unique class HDInitStates(DialinEnum): - POST_STATE_START = 0 # Start initialize & POST mode state - POST_STATE_FW_INTEGRITY = 1 # Run firmware integrity test state - POST_STATE_AC = 2 # Check HD has AC power - power off if not - POST_STATE_WATCHDOG = 3 # Run watchdog test state - POST_STATE_SAFETY_SHUTDOWN = 4 # Run safety shutdown test state - POST_STATE_RTC = 5 # Run RTC test state - POST_STATE_NVDATAMGMT = 6 # Run NV Data Mgmt. test state - POST_STATE_BLOOD_FLOW = 7 # Run blood flow test state - POST_STATE_DIALYSATE_INLET_FLOW = 8 # Run dialysate inlet flow test state - POST_STATE_DIALYSATE_OUTLET_FLOW = 9 # Run dialysate outlet flow test state - POST_STATE_BLOOD_LEAK = 10 # Run blood leak sensor test state - POST_STATE_VALVES = 11 # Run valves test state - POST_STATE_SYRINGE_PUMP = 12 # Run syringe pump test state - POST_STATE_PRES_OCCL = 13 # Run pressure occlusion state - POST_STATE_ALARM_AUDIO = 14 # Run alarm audio test state - POST_STATE_ALARM_LAMP = 15 # Run alarm lamp test state - POST_STATE_ACCELEROMETER = 16 # Run Accelerometer test state - POST_STATE_TEMPERATURES = 17 # Run temperatures POST state - POST_STATE_FANS = 18 # Run fans POST state - POST_STATE_STUCK_BUTTON = 19 # Run stuck button test state - POST_STATE_UI_POST = 20 # Check whether UI passed its POST tests - POST_STATE_FW_COMPATIBILITY = 21 # Run firmware compatibility test state - POST_STATE_FPGA = 22 # Run FPGA test state - POST_STATE_COMPLETED = 23 # POST self-tests completed state - POST_STATE_FAILED = 24 # POST self-tests failed state - NUM_OF_POST_STATES = 25 # Number of initialize & POST mode states + POST_STATE_START = 0 # Start initialize & POST mode state + POST_STATE_FW_INTEGRITY = 1 # Run firmware integrity test state + POST_STATE_RTC = 2 # Run RTC test state + POST_STATE_NVDATAMGMT = 3 # Run NV Data Mgmt. test state + POST_STATE_WATCHDOG = 4 # Run watchdog test state + POST_STATE_SAFETY_SHUTDOWN = 5 # Run safety shutdown test state + POST_STATE_BLOOD_FLOW = 6 # Run blood flow test state + POST_STATE_DIALYSATE_INLET_FLOW = 7 # Run dialysate inlet flow test state + POST_STATE_DIALYSATE_OUTLET_FLOW = 8 # Run dialysate outlet flow test state + POST_STATE_BLOOD_LEAK = 9 # Run blood leak sensor test state + POST_STATE_VALVES = 10 # Run valves test state + POST_STATE_SYRINGE_PUMP = 11 # Run syringe pump test state + POST_STATE_PRES_OCCL = 12 # Run pressure occlusion state + POST_STATE_ALARM_AUDIO = 13 # Run alarm audio test state + POST_STATE_ALARM_LAMP = 14 # Run alarm lamp test state + POST_STATE_ACCELEROMETER = 15 # Run Accelerometer test state + POST_STATE_TEMPERATURES = 16 # Run temperatures POST state + POST_STATE_FANS = 17 # Run fans POST state + POST_STATE_STUCK_BUTTON = 18 # Run stuck button test state + POST_STATE_UI_POST = 19 # Check whether UI passed its POST tests + POST_STATE_FW_COMPATIBILITY = 20 # Run firmware compatibility test state + POST_STATE_FPGA = 21 # Run FPGA test state + POST_STATE_COMPLETED = 22 # POST self-tests completed state + POST_STATE_FAILED = 23 # POST self-tests failed state + NUM_OF_POST_STATES = 24 # Number of initialize & POST mode states @unique class PreTreatmentSubModes(DialinEnum):